Image Quality and Radiation Dose of Pulmonary CT Angiography Performed Using 100 and 120 kVp

被引:41
作者
Fanous, Randy [1 ]
Kashani, Hany [1 ]
Jimenez, Laura [1 ]
Murphy, Grainne [1 ]
Paul, Narinder S. [1 ]
机构
[1] Toronto Gen Hosp, Univ Hlth Network, Dept Med Imaging, Toronto, ON M5G 2N2, Canada
关键词
CT angiography; pulmonary embolism; radiation dose; COMPUTED-TOMOGRAPHY ANGIOGRAPHY; MULTIDETECTOR ROW CT; CHEST CT; VASCULAR ENHANCEMENT; FEMALE BREAST; STANDARD; EMBOLISM; REDUCTION; EXPOSURE; OPTIMIZATION;
D O I
10.2214/AJR.11.8208
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
OBJECTIVE. The objective of our study was to compare image quality and radiation dose of pulmonary CT angiography (CTA) performed in the same patient cohort using tube potentials of 100 and 120 kVp. MATERIALS AND METHODS. The study group for this retrospective study was 32 patients (22 women, 10 men) with a mean age of 57 years (age range, 28-83 years; body weight < 100 kg). Patients underwent pulmonary CTA studies performed using 120 and 100 kVp while other scanning parameters were kept constant. Two observers measured image signal and image noise, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR), and SNR dose and CNR dose. Two additional observers performed qualitative image quality analysis using a 5-point grading scale (5 = excellent). RESULTS. The reduction in tube potential caused image signal to increase by 29% (p < 0.0001), image noise to increase by 68% (p < 0.0001), CNR dose to decrease by 0.8% (p = 0.91) and SNR to decrease by 24% (p = 0.0002) and CNR by 20% (p = 0.0019). Radiation dose (dose-length product) was decreased by 37% to 379.26 mGy x cm at 100 kVp from 604.46 mGy x cm at 120 kVp (p < 0.0001). The median pulmonary arteries image quality scores for observers 1 and 2, respectively, were as follows at 100 kVp: main, 5 and 5; lobar, 5 and 4.5; and segmental, 5 and 4. At 120 kVp, the median image quality scores for observers 1 and 2 were as follows: main, 5 and 5; lobar, 5 and 5; segmental, 4 and 4. A Wilcoxon test analysis indicated no significant difference in image quality between the studies (main, p = 0.59; lobar, p = 0.88; segmental, p = 0.79). CONCLUSION. Pulmonary CTA can be performed using a tube potential of 100 kVp in patients who weigh less than 100 kg (220 lb). Reducing the tube potential from 120 to 100 kVp results in a 37% reduction in radiation dose without a significant impact on diagnostic image quality.
引用
收藏
页码:990 / 996
页数:7
相关论文
共 35 条
[1]
[Anonymous], 2004, EUROPEAN GUIDELINES
[2]
Current concepts - Computed tomography - An increasing source of radiation exposure [J].
Brenner, David J. ;
Hall, Eric J. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) :2277-2284
[3]
Bushberg JT, 2002, ESSENTIAL PHYS MED I, P255
[4]
Radiology resident interpretation of on-call CT pulmonary angiograms [J].
Cervini, Patrick ;
Bell, Chaim M. ;
Roberts, Heidi C. ;
Provost, Yves L. ;
Chung, Tae-Bong ;
Paul, Narinder S. .
ACADEMIC RADIOLOGY, 2008, 15 (05) :556-562
[5]
Minimizing radiation dose for pediatric body applications of single-detector helical CT: Strategies at a large children's hospital [J].
Donnelly, LF ;
Emery, KH ;
Brody, AS ;
Laor, T ;
Gylys-Morin, VM ;
Anton, CG ;
Thomas, SR ;
Frush, DP .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 176 (02) :303-306
[6]
Radiation dose to patients from cardiac diagnostic imaging [J].
Einstein, Andrew J. ;
Moser, Kevin W. ;
Thompson, Randall C. ;
Cerqueira, Manuel D. ;
Henzlova, Milena J. .
CIRCULATION, 2007, 116 (11) :1290-1305
[7]
Evaluation of Peripheral Pulmonary Arteries at 80 kV and at 140 kV: Dual-Energy Computed Tomography Assessment in 51 Patients [J].
Gorgos, Andrei ;
Remy-Jardin, Martine ;
Duhamel, Alain ;
Faivre, Jean-Baptiste ;
Tacelli, Nunzia ;
Delannoy, Valerie ;
Remy, Jacques .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2009, 33 (06) :981-986
[8]
THE EFFECT OF MAS VARIATION UPON COMPUTED-TOMOGRAPHY IMAGE QUALITY AS EVALUATED BY INVIVO AND INVITRO STUDIES [J].
HAAGA, JR ;
MIRALDI, F ;
MACINTYRE, W ;
LIPUMA, JP ;
BRYAN, PJ ;
WIESEN, E .
RADIOLOGY, 1981, 138 (02) :449-454
[9]
Multi-detector row CT: Radiation dose characteristics [J].
Hamberg, LM ;
Rhea, JT ;
Hunter, GJ ;
Thrall, JH .
RADIOLOGY, 2003, 226 (03) :762-772
[10]
Image quality and radiation exposure at pulmonary CT angiography with 100-or 120-kVp protocol: prospective randomized study [J].
Heyer, Christoph M. ;
Mohr, Patrick S. ;
Lemburg, Stefan P. ;
Peters, Soeren A. ;
Nicolas, Volkmar .
RADIOLOGY, 2007, 245 (02) :577-583