Microwave digestion and ultrasonic nebulization for determination of boron in animal tissues by inductively coupled plasma atomic emission spectrometry with internal standardization and addition of mannitol

被引:57
作者
Sun, DH [1 ]
Waters, JK [1 ]
Mawhinney, TP [1 ]
机构
[1] UNIV MISSOURI,EXPT STN,CHEM LABS,COLUMBIA,MO 65211
关键词
boron; ultrasonic nebulizer; microwave digestion; mannitol; inductively coupled plasma; internal standardization;
D O I
10.1039/a606283k
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Determinations of boron in animal tissues were performed by inductively coupled plasma atomic emission spectrometry (ICP-AES) with an ultrasonic nebulizer. The method is based on the microwave digestion of tissues with HNO3-H2O2 and the use of mannitol as a modifier. It was found that mannitol can significantly enhance the analytical sensitivity, improve the precision and minimize the memory effect for the determination of boron. Acid effect and matrix effects were controlled by using the Be I 234.861 nm line as the internal standard. Six animal tissue samples, including two NIST standard reference materials (Oyster Tissue and Bovine Liver), were analyzed to test the reliability of the method. A limit of detection (3 sigma) of 0.7 mu g l(-1) and recoveries of added boron from selected matrices between 95.0 and 100.2% were obtained.
引用
收藏
页码:675 / 679
页数:5
相关论文
共 21 条
[1]   INTERFERENCE FROM EASILY IONIZABLE ELEMENT MATRICES IN INDUCTIVELY COUPLED PLASMA EMISSION-SPECTROMETRY - A SPATIAL STUDY [J].
BLADES, MW ;
HORLICK, G .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1981, 36 (09) :881-900
[2]  
BOUMAN PWJ, 1987, INDUCTIVELY COUPLED, pCH7
[3]  
Cunniff P. A., 1995, Official methods of analysis of AOAC international, V16th
[4]  
DENTON MB, 1990, SAMPLE INTRO ATOMIC, pCH4
[5]   IMPROVED BORON DETERMINATION IN BIOLOGICAL-MATERIAL BY INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY [J].
EVANS, S ;
KRAHENBUHL, U .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (11) :1249-1253
[6]   MICROWAVE DIGESTION PREPARATION AND ICP DETERMINATION OF BORON IN HUMAN PLASMA [J].
FERRANDO, AA ;
GREEN, NR ;
BARNES, KW ;
WOODWARD, B .
BIOLOGICAL TRACE ELEMENT RESEARCH, 1993, 37 (01) :17-25
[7]   ULTRASONIC NEBULIZATION IN INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
GALLI, P ;
ODDO, N .
MICROCHEMICAL JOURNAL, 1992, 46 (03) :327-334
[8]   SUPPRESSION OF BORON VOLATILIZATION FROM A HYDROFLUORIC-ACID SOLUTION USING A BORON-MANNITOL COMPLEX [J].
ISHIKAWA, T ;
NAKAMURA, E .
ANALYTICAL CHEMISTRY, 1990, 62 (23) :2612-2616
[9]  
Kingston H.M., 1988, INTRO MICROWAVE SAMP
[10]   APPLICATIONS OF MICROWAVE DIGESTION TECHNIQUE FOR ELEMENTAL ANALYSES [J].
KUSS, HM .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1992, 343 (9-10) :788-793