Radiation dose in dental radiology

被引:129
作者
Cohnen, M
Kemper, J
Möbes, O
Pawelzik, J
Mödder, U
机构
[1] Univ Dusseldorf, Inst Diagnost Radiol, D-40225 Dusseldorf, Germany
[2] Univ Dusseldorf, Dept Oral Surg, Westdeutsch Kieferklin, D-4000 Dusseldorf, Germany
关键词
dental radiology; radiation exposure; dose reduction; image noise;
D O I
10.1007/s003300100928
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study was to compare radiation exposure in panoramic radiography (PR), dental CT and digital volume tomography (DVT). An anthropomorphic Alderson-Rando phantom and two anatomical head phantoms with thermoluminescent dosimeters fixed at appropriate locations were exposed as in a dental examination. In PR and DVT standard parameters were used while variables in CT included mA. pitch. and rotation time. Image noise was assessed in dental CT and DVT. Radiation doses to the skin and internal organs within the primary beam and resulting from scatter radiation were measured and expressed as maximum doses in mGy. For PR, DVT, and CT these maximum doses were 0.65, 4.2. and 23 mGy. In dose-reduced CT protocols, radiation doses ranged from 10.9 to 6.1 mGy. Effective doses calculated on this basis showed values below 0.1 mSv for PR, DVT, and dose-reduced CT. Image noise was similar in DVT and low-dose CT. As radiation exposure and image noise of DVT is similar to low-dose CT this imaging technique cannot be recommended as a general alternative to replace PR in dental radiology.
引用
收藏
页码:634 / 637
页数:4
相关论文
共 10 条
[1]   DENTAL IMPLANT TREATMENT PLANNING WITH REFORMATTED COMPUTED-TOMOGRAPHY [J].
BESIMO, C ;
LAMBRECHT, JT ;
NIDECKER, A .
DENTOMAXILLOFACIAL RADIOLOGY, 1995, 24 (04) :264-267
[2]   In vivo, thyroid and lens surface exposure with spiral and conventional computed tomography in dental implant radiography [J].
Bianchi, J ;
Goggins, W ;
Rudolph, M .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 2000, 90 (02) :249-253
[3]   An approach to the diagnostic imaging of jaw lesions, dental implants, and the temporomandibular joint [J].
DelBalso, AM .
RADIOLOGIC CLINICS OF NORTH AMERICA, 1998, 36 (05) :855-+
[4]  
HASSFELD S, 1998, DTSCH Z MUND KIEFER, V2, P188
[5]  
KORNAS M, 1998, DTSCH ZAHNARZTL Z, V53, P120
[6]  
Kraut R A, 1998, Implant Dent, V7, P19, DOI 10.1097/00008505-199804000-00002
[7]  
Lenglinger FX, 1999, RADIOLOGE, V39, P1027, DOI 10.1007/s001170050598
[8]   A new volumetric CT machine for dental imaging based on the cone-beam technique: preliminary results [J].
Mozzo, P ;
Procacci, C ;
Tacconi, A ;
Martini, PT ;
Andreis, IAB .
EUROPEAN RADIOLOGY, 1998, 8 (09) :1558-1564
[9]   Low-dose dental CT. [J].
Rustemeyer, P ;
Streubühr, U ;
Hohn, HP ;
Rustemeyer, R ;
Eich, HT ;
John-Mikolajewski, V ;
Müller, RD .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1999, 171 (02) :130-135
[10]  
Schorn C, 1999, ROFO-FORTSCHR RONTG, V170, P137