TrueFISP: Assessment of accuracy for measurement of left ventricular mass in an animal model

被引:70
作者
Fieno, DS [1 ]
Jaffe, WC [1 ]
Simonetti, OP [1 ]
Judd, RM [1 ]
Finn, JP [1 ]
机构
[1] Northwestern Univ, Sch Med, Dept Radiol, Chicago, IL 60611 USA
关键词
ventricular mass; fast imaging; FISP; accuracy; animals;
D O I
10.1002/jmri.10107
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To test the accuracy of a high performance true fast imaging with steady-state precession (TrueFISP) pulse sequence for the assessment of left ventricular (LV) mass in a large animal model on 1.5-T scanners. Materials and Methods: We imaged dogs (N = 10) on a clinical 1.5-T clinical scanner using electrocardiogram (ECG)-gated TrueFISP. In all animals, contiguous segmented k-space cine images were acquired from base to apex (in-plane resolution 1 X 1 mm(2), slice thickness 5 mm, TR = 4.8 msec, TE = 1.6 msec) during repeated breathholds. In nine of the 10 animals, single-shot images gated to end-diastole were also acquired from base to apex in a single breath-hold (in-plane resolution 1 X 1 mm(2), slice thickness 5 mm, TR = 3,2 msec, TE = 1.6 msec). After imaging, animals were killed, the left ventricle was isolated, and the true mass of the left ventricle (free wall and septum) was determined. Independently, two observers blinded to the post-mortem results computed LV masses based on analysis of the magnetic resonance (MR) images. Results: Comparison of the computed LV mass using TrueFISP to the actual mass showed excellent agreement. Cinesystole was the most accurate technique (mass = 98.6% +/- 4.5% actual, bias = 1.2 +/- 3.4 g) followed by cine-diastole (mass = 97.9% +/- 5.3% actual, bias = 1.8 +/- 4.1 g) and single shot (mass = 94.7% +/- 7.9% actual, bias = 4.2 +/- 6.3 g). Inter- and intra-observer variabilities were low (5.8% +/- 7.1% and 0.4% +/- 4.8%, respectively). Conclusion: We conclude that TrueFISP imaging is an accurate, rapid method to determine ventricular mass. In single-shot mode, TrueFISP requires only one breath-hold to estimate the mass of the heart within 6% of the actual value, whereas the segmented k-space implementation measured LV mass to within 3% of the true value. (C) 2002 Wiley-Liss, Inc.
引用
收藏
页码:526 / 531
页数:6
相关论文
共 28 条
  • [1] *AM HEART ASS, 1985, CIRCULATION, V71, pA849
  • [2] Measuring agreement in method comparison studies
    Bland, JM
    Altman, DG
    [J]. STATISTICAL METHODS IN MEDICAL RESEARCH, 1999, 8 (02) : 135 - 160
  • [3] COMPUTER-ANALYSIS OF LEFT-VENTRICULAR DYNAMIC GEOMETRY IN MAN
    BOVE, AA
    KREULEN, TH
    SPANN, JF
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1978, 41 (07) : 1239 - 1248
  • [4] Bundy J, 1999, P 7 ANN M ISMRM PHIL, P1282
  • [5] DETERMINATION OF LEFT-VENTRICULAR VOLUME AND MASS WITH USE OF BIPHASIC SPIN-ECHO MR IMAGING - COMPARISON WITH CINE MR
    CAPUTO, GR
    SUZUKI, JI
    KONDO, C
    CHO, H
    QUAIFE, RA
    HIGGINS, CB
    PARKER, DL
    [J]. RADIOLOGY, 1990, 177 (03) : 773 - 777
  • [6] MEASUREMENT OF CANINE LEFT-VENTRICULAR MASS BY USING MR IMAGING
    CAPUTO, GR
    TSCHOLAKOFF, D
    SECHTEM, U
    HIGGINS, CB
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1987, 148 (01) : 33 - 38
  • [7] STEADY-STATE FREE PRECESSION IN NUCLEAR MAGNETIC RESONANCE
    CARR, HY
    [J]. PHYSICAL REVIEW, 1958, 112 (05): : 1693 - 1701
  • [8] Cine MR angiography of the heart with segmented true fast imaging with steady-state precession
    Carr, JC
    Simonetti, O
    Bundy, J
    Li, DB
    Pereles, S
    Finn, JP
    [J]. RADIOLOGY, 2001, 219 (03) : 828 - 834
  • [9] ECHOCARDIOGRAPHICALLY DETERMINED LEFT-VENTRICULAR MASS INDEX IN NORMAL-CHILDREN, ADOLESCENTS AND YOUNG-ADULTS
    DANIELS, SR
    MEYER, RA
    LIANG, YC
    BOVE, KE
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1988, 12 (03) : 703 - 708
  • [10] Left ventricular remodelling: common process in patients with different primary myocardial disorders
    Florea, VG
    Mareyev, VY
    Samko, AN
    Orlova, IA
    Coats, AJS
    Belenkov, YN
    [J]. INTERNATIONAL JOURNAL OF CARDIOLOGY, 1999, 68 (03) : 281 - 287