Coronary thermodilution to assess flow reserve - Validation in humans

被引:254
作者
Pijls, NHJ
De Bruyne, B
Smith, L
Aarnoudse, W
Barbato, E
Bartunek, J
Bech, GJW
Van de Vosse, F
机构
[1] Catharina Hosp, Dept Cardiol, NL-5602 ZA Eindhoven, Netherlands
[2] Ctr Cardiovasc, Dept Cardiol, Aalst, Belgium
[3] Radi Med Syst, Uppsala, Sweden
[4] Eindhoven Univ Technol, Dept Biomed Engn, NL-5600 MB Eindhoven, Netherlands
关键词
blood flow; stenosis; arteries; microcirculation; coronary artery disease;
D O I
10.1161/01.CIR.0000017199.09457.3D
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background-Guide wire-based simultaneous measurement of fractional flow reserve (FFR) and coronary flow reserve (CFR) is important to understand microvascular disease of the heart. The aim of this study was to investigate the feasibility of simultaneous measurement of FFR and CFR by one pressure-temperature sensor-tipped guide wire with the use of coronary thermodilution and to compare CFR by thermodilution (CFRthermo) with simultaneously measured Doppler CFR (CFRDoppl) Methods and Results-In 103 coronary arteries in 50 patients, a pressure-temperature sensor-tipped 0.014-inch floppy guide wire and a 0.014-inch Doppler guide wire were introduced. Both normal vessels and a wide range of stenotic vessels were included. With 3 mL of saline at room temperature used as an indicator, by hand-injection, thermodilution curves in the coronary artery were obtained in triplicate, both at baseline and at intravenous adenosine-induced maximum hyperemia. After adequate curve-fitting CFRthermo was calculated from the ratio of inverse mean transit times and compared with CFRDoppl calculated by velocities at hyperemia and baseline. Adequate sets of thermodilution curves and corresponding CFRthermo could be obtained in 87% of the arteries versus 91 % for Doppler CFR and 100% for FFR. CFRthermo correlated fairly well to CFRDoppl (CFRthermo =0.84 CFRDoppl+0.17; r=0.80; P<0.001), although individual differences of >20% between both indexes were seen in a quarter of all arteries. Conclusions-This study shows the feasibility of simultaneous measurement of FFR (by coronary pressure) and CFR (by coronary thermodilution) in humans by one single guide wire in a practical and straightforward way and will facilitate assessment of microvascular disease.
引用
收藏
页码:2482 / 2486
页数:5
相关论文
共 22 条
[1]   CORONARY-ARTERY FLOW MONITORING FOLLOWING CORONARY INTERVENTIONS [J].
ANDERSON, HV ;
KIRKEEIDE, RL ;
STUART, Y ;
SMALLING, RW ;
HEIBIG, J ;
WILLERSON, JT .
AMERICAN JOURNAL OF CARDIOLOGY, 1993, 71 (14) :D62-D69
[2]   Usefulness of fractional flow reserve to predict clinical outcome after balloon angioplasty [J].
Bech, GJW ;
Pijls, NHJ ;
De Bruyne, B ;
Peels, KH ;
Michels, HR ;
Bonnier, HJRM ;
Koolen, JJ .
CIRCULATION, 1999, 99 (07) :883-888
[3]  
CASCALLS W, 1996, LANCET, V347, P1447
[4]   Pressure-derived fractional flow reserve to assess serial epicardial stenoses - Theoretical basis and animal validation [J].
De Bruyne, B ;
Pijls, NHJ ;
Heyndrickx, GR ;
Hodeige, D ;
Kirkeeide, R ;
Gould, KL .
CIRCULATION, 2000, 101 (15) :1840-1847
[5]   Coronary thermodilution to assess flow reserve - Experimental validation [J].
De Bruyne, B ;
Pijls, NHJ ;
Smith, L ;
Wievegg, M ;
Heyndrickx, GR .
CIRCULATION, 2001, 104 (17) :2003-2006
[6]   Abnormal epicardial coronary resistance in patients with diffuse atherosclerosis but "normal" coronary angiography [J].
De Bruyne, B ;
Hersbach, F ;
Pijls, NHJ ;
Bartunek, J ;
Bech, JW ;
Heyndrickx, GR ;
Gould, KL ;
Wijns, W .
CIRCULATION, 2001, 104 (20) :2401-2406
[7]   Fractional flow reserve in patients with prior myocardial infarction [J].
De Bruyne, B ;
Pijls, NHJ ;
Bartunek, J ;
Kulecki, K ;
Bech, JW ;
De Winter, H ;
Van Crombrugge, P ;
Heyndrickx, GR ;
Wijns, W .
CIRCULATION, 2001, 104 (02) :157-162
[8]   Simultaneous coronary pressure and flow velocity measurements in humans - Feasibility, reproducibility, and hemodynamic dependence of coronary flow velocity reserve, hyperemic flow versus pressure slope index, and fractional flow reserve [J].
deBruyne, B ;
Bartunek, J ;
Sys, SU ;
Pijls, NHJ ;
Heyndrickx, GR ;
Wijns, W .
CIRCULATION, 1996, 94 (08) :1842-1849
[9]   SLOPE OF THE INSTANTANEOUS HYPEREMIC DIASTOLIC CORONARY FLOW VELOCITY-PRESSURE RELATION - A NEW INDEX FOR ASSESSMENT OF THE PHYSIOLOGICAL SIGNIFICANCE OF CORONARY STENOSIS IN HUMANS [J].
DIMARIO, C ;
KRAMS, R ;
GIL, R ;
SERRUYS, PW .
CIRCULATION, 1994, 90 (03) :1215-1224
[10]   Comparison of quantitative coronary angiography, intravascular ultrasound, and coronary pressure measurement to assess optimum stent deployment [J].
Hanekamp, CEE ;
Koolen, JJ ;
Pijls, NHJ ;
Michels, HR ;
Bonnier, HJRM .
CIRCULATION, 1999, 99 (08) :1015-1021