Initial stability of two PLIF-techniques.: A biomechanical in-vitro-comparison

被引:9
作者
Pitzen, T [1 ]
Caspar, W [1 ]
Matthis, D [1 ]
Müller-Storz, H [1 ]
König, J [1 ]
Georg, T [1 ]
Wurm, EM [1 ]
Steudel, WI [1 ]
机构
[1] Univ Saarland, Neurochirurg Klin, D-66421 Homburg, Germany
来源
ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE | 1999年 / 137卷 / 03期
关键词
spine; posterior interbody fusion; biomechanics; in-vitro-study; crossover-study;
D O I
10.1055/s-2008-1037396
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective: The purpose of this biomechanical in-vitro-study was to compare two different PLIF-techniques with two types of implants on human lumbar spine: PLIF with threaded cages, (Bagby and Kuslich, Spinetech, Minneapolis, USA) and PLIF with the Moss-Miami-implants, (DePuy International Limited, Leeds, Great Britain). Methods: Six cadaveric human lumbar spine segments L2-5 were explanted, frozen at -20 degrees C and thawed before preparation. They were cut in two parts by discectomie and arthrotomie L3/4, so six specimen L2/3 and six specimen L4/5 were obtained and used in a crossover-trial. Analysis included testing in a tension-torsion-machine under axial compression with 600 N, rotation (left-right) with 25 Nm and shearing forces with 250 N without preload. This was first done in the intact and then in the fused specimen. Results: Stiffness before treatment was comparable in both groups irrespective of location. Posttreatment stiffness was higher with MOSS-MIAMI-implants as compared to PLIF with BAK-cages. Average relative superiority (and 95%-confidence intervall) were 1,98 (1,01-3,69) for compression, 2,30 (0,85-6,24) for rotation and 1,73 (0,78-3,84) for shearing. Statistical comparison of log posttreatment: stiffness was significant for compression but not for rotation and shearing (2-sided independent crossover t-test). Conclusion: This biomechanical in-vitro-study demonstrates the higher initial stability of PLIF with titanium surgical mesh and posterior instrumentation when compared to PLIF with threaded cages alone.
引用
收藏
页码:214 / 218
页数:5
相关论文
共 21 条
[11]   Strength and stability of posterior lumbar interbody fusion - Comparison of titanium fiber mesh implant and tricortical bone graft [J].
Hoshijima, K ;
Nightingale, RW ;
Yu, JR ;
Richardson, WJ ;
Harper, KD ;
Yamamoto, H ;
Myers, BS .
SPINE, 1997, 22 (11) :1181-1188
[12]  
IMPLACITO AD, 1997, 3 STRYK INV S 2 3 OK
[13]   The Bagby and Kuslich method of lumbar interbody fusion - History, techniques, and 2-year follow-up results of a United States prospective, multicenter trial [J].
Kuslich, SD ;
Ulstrom, CL ;
Griffith, SL ;
Ahern, JW ;
Dowdle, JD .
SPINE, 1998, 23 (11) :1267-1278
[14]  
LIN PM, 1985, CLIN ORTHOP RELAT R, P90
[15]   Threaded titanium cages for lumbar interbody fusions [J].
Ray, CD .
SPINE, 1997, 22 (06) :667-679
[16]   Threaded fusion cages for lumbar interbody fusions - An economic comparison with 360 degrees fusions [J].
Ray, CD .
SPINE, 1997, 22 (06) :681-685
[17]  
Tullberg T, 1996, Eur Spine J, V5, P178, DOI 10.1007/BF00395510
[18]  
WHITESIDES TE, 1977, CLIN ORTHOP RELAT R, V128, P78
[19]  
WILKE HJ, 1998, EUR SPINE J, P148
[20]  
ZDEBLICK TA, 1993, SPINE, V18, P513