Fat suppressed MRI of articular cartilage with a spatial-spectral excitation pulse

被引:59
作者
Hardy, PA [1 ]
Recht, MP [1 ]
Piraino, DW [1 ]
机构
[1] Cleveland Clin Fdn, Div Biomed Engn, Cleveland, OH 44195 USA
来源
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING | 1998年 / 8卷 / 06期
关键词
articular cartilage; RF pulses; three-dimensional imaging;
D O I
10.1002/jmri.1880080615
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We developed a three-dimensional, gradient-recalled-echo imaging technique that incorporates a short-duration spatial-spectral excitation pulse from the family of binomial pulses. Binomial pulses of different orders were tested on phantoms and on normal volunteers to find the composite pulse that produced in the shortest duration the most reliable fat suppression. Composite pulses employing unipolar slice-selective gradients with explicit rewinder gradients between each radio-frequency (RF) pulse were compared with composite RF pulses employing alternating-polarity, slice-select gradients. The advantage of the sequences using the unipolar gradients is improved fat suppression. Images of the knees of volunteers produced with the composite RF pulse have contrast between fat and articular cartilage equivalent to that on images created by the gradient-recalled-echo imaging technique employing a conventional chemsat pulse. The optimum RF pulse consisted of three amplitude- and phase-modulated pulses combined with unipolar slice-select gradients.
引用
收藏
页码:1279 / 1287
页数:9
相关论文
共 22 条
  • [21] THOMASSON DM, 1996, P 4 ANN M INT SOC MA, P1462