Wireless intra-ocular pressure monitoring system integrated into an artificial lens

被引:18
作者
Eggers, T [1 ]
Draeger, J [1 ]
Hille, K [1 ]
Marschner, C [1 ]
Stegmaier, P [1 ]
Binder, J [1 ]
Laur, R [1 ]
机构
[1] Campus Micro Technol GMBH, D-28359 Bremen, Germany
来源
1ST ANNUAL INTERNATIONAL IEEE-EMBS SPECIAL TOPIC CONFERENCE ON MICROTECHNOLOGIES IN MEDICINE & BIOLOGY, PROCEEDINGS | 2000年
关键词
D O I
10.1109/MMB.2000.893827
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
This paper reports the concept and first results of a hybrid integrated telemetric pressure measuring system for continuous monitoring of the intra-ocular pressure. The system is designed for glaucoma differential diagnosis and follow up. The measuring system is highly miniaturized using advanced flip-chip mounting technology to fulfil the spatial requirements of integration into the hinge of an intra-ocular lens without the necessity of a monolithically integrated sensor. The surface micro-machined pressure sensor is used in an integrated hybrid circuit comprising the sensor itself and a dedicated read-out circuit in standard CMOS technology. First experimental results of the system will be presented. The concept of the telemetric data transmission, mounting and assembly, the coating and EMC problem will be discussed briefly.
引用
收藏
页码:466 / 469
页数:4
相关论文
共 13 条
[1]   PASSIVE SILICON TRANSENSOR INTENDED FOR BIOMEDICAL, REMOTE PRESSURE MONITORING [J].
BACKLUND, Y ;
ROSENGREN, L ;
HOK, B ;
SVEDBERGH, B .
SENSORS AND ACTUATORS A-PHYSICAL, 1990, 21 (1-3) :58-61
[2]  
CICHOCKI A, 1989, EUR C CIRC THEOR DES
[4]   National cataract surgery survey 1997-8. A report of the results of the clinical outcomes [J].
Desai, P ;
Minassian, DC ;
Reidy, A .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1999, 83 (12) :1336-1340
[5]  
DUDENBOSTEL D, 1997, TRANSDUCERS 97
[6]  
EGGERS T, 2000, ADV HYBRID INTEGRATE
[7]   A high-efficiency power and data transmission system for biomedical implanted electronic devices [J].
Hamici, Z ;
Itti, R ;
Champier, J .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1996, 7 (02) :192-201
[8]  
HIEROLD C, 1998, IMPLANTABLE LOW POWE
[9]  
McLaren JW, 1996, INVEST OPHTH VIS SCI, V37, P966
[10]  
Repacholi MH, 1998, BIOELECTROMAGNETICS, V19, P1, DOI 10.1002/(SICI)1521-186X(1998)19:1<1::AID-BEM1>3.0.CO