Background: A recent report linked shelf-aging of unicondylar polyethylene tibial components with accelerated fatigue wear and rapid failure. We retrospectively reviewed our experiences with another unicondylar knee system to investigate the relationship between the shelf age of the polyethylene components and clinical outcome. Methods: One hundred Single Compartment Replacement unicompartmental knee arthroplasties were performed with cement between 1990 and 1996. The median shelf age for the polyethylene inserts was 1.7 years. At the time of the review, four knees had been lost to follow-up, sixteen were in patients who had died, and nineteen had been revised. Sixty-one knee replacements remained in situ at a mean and standard deviation of 8 2 years postoperatively. Results: With revision as the end point, Kaplan-Meier survivorship analysis demonstrated a six-year rate of survival of 96% when the shelf age of the insert was less than the median shelf age but only 71% when the median shelf age was exceeded. Fatigue wear was identified on all retrieved components except for two that had shelf ages under one year and had been revised within three years. At four to six years of clinical follow-up, the group of knees in which the median shelf age had been exceeded also had poorer Knee Society knee scores and function scores. Conclusions: Aging on the shelf accelerated fatigue failure of polyethylene inserts sterilized with gamma irradiation in air and compromised our intermediate-term clinical outcomes with this unicondylar knee replacement system. Attention to the relationship between the shelf age and the clinical performance of the polyethylene component continues to be warranted with unicondylar knee arthroplasty implants. Level of Evidence: Therapeutic study, Level III-2 (retrospective cohort study). See Instructions to Authors for a complete description of levels of evidence.
引用
收藏
页码:763 / 769
页数:7
相关论文
共 20 条
[1]  
BARTLEY RE, 1994, CLIN ORTHOP RELAT R, P18
[2]  
BERGENUDD H, 1995, J ARTHROPLASTY, V10, P8
[3]  
Berger RA, 1999, CLIN ORTHOP RELAT R, P50
[4]   POLYETHYLENE WEAR IN UNICONDYLAR KNEE PROSTHESES - 106 RETRIEVED MARMOR, PCA, AND ST-GEORG TIBIAL COMPONENTS COMPARED [J].
BLUNN, GW ;
JOSHI, AB ;
LILLEY, PA ;
ENGELBRECHT, E ;
RYD, L ;
LIDGREN, L ;
HARDINGE, K ;
NIEDER, E ;
WALKER, PS .
ACTA ORTHOPAEDICA SCANDINAVICA, 1992, 63 (03) :247-255
[5]  
Bohl JR, 1999, CLIN ORTHOP RELAT R, P28
[6]  
Currier BH, 1997, CLIN ORTHOP RELAT R, P111
[7]   Questionnaire on the perceptions of patients about total knee replacement [J].
Dawson, J ;
Fitzpatrick, R ;
Murray, D ;
Carr, A .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1998, 80B (01) :63-69
[8]  
INSALL J, 1976, CLIN ORTHOP RELAT R, P83
[9]  
INSALL JN, 1989, CLIN ORTHOP RELAT R, P13
[10]   COMPARISON OF 4 MODELS OF TOTAL KNEE-REPLACEMENT PROSTHESES [J].
INSALL, JN ;
RANAWAT, CS ;
AGLIETTI, P ;
SHINE, J .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1976, 58 (06) :754-765