POLYETHYLENE WEAR IN UNICONDYLAR KNEE PROSTHESES - 106 RETRIEVED MARMOR, PCA, AND ST-GEORG TIBIAL COMPONENTS COMPARED

被引:77
作者
BLUNN, GW
JOSHI, AB
LILLEY, PA
ENGELBRECHT, E
RYD, L
LIDGREN, L
HARDINGE, K
NIEDER, E
WALKER, PS
机构
[1] WRIGHTINGTON HOSP, CTR HIP SURG, WIGAN, ENGLAND
[2] UNIV LUND HOSP, DEPT ORTHOPED, S-22185 LUND, SWEDEN
[3] ENDOKLIN, HAMBURG, GERMANY
来源
ACTA ORTHOPAEDICA SCANDINAVICA | 1992年 / 63卷 / 03期
关键词
D O I
10.3109/17453679209154776
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
106 unicondylar knee replacement tibial components were retrieved and analyzed for the amount and type of polyethylene wear. Three different designs were retrieved which had essentially the same femorotibial conformity. Each design showed a characteristic failure pattern. The polyethylene of PCA tibial components showed serious delamination after only short durations, as a result of heat pressing. St Georg sledge prostheses showed some delamination after 4 years' duration due to sub-surface cracks which were initiated by fusion defects in the polyethylene; metal backing of the components did not affect delamination of this prosthesis. The Marmor designs showed the least wear, with shiny depressions and surface pitting; no delamination was observed in the Marmor prosthesis. Molecular weight determination by gel permeation chromatography and analysis of crystallinity using Fourier transformation infra-red spectroscopy demonstrated that St Georg polyethylene had higher molecular weight and crystallinity than Marmor polyethylene. In some of the components investigated, crystallinity and molecular weight of the polyethylene were reduced under the wear track when compared with the unworn polyethylene. Since fusion defects may cause delamination of polyethylene we urge manufacturers to reduce the number of such defects.
引用
收藏
页码:247 / 255
页数:9
相关论文
共 23 条
  • [1] THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT
    BARTEL, DL
    BICKNELL, VL
    WRIGHT, TM
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) : 1041 - 1051
  • [2] BLOEBAUM RD, 1991, CLIN ORTHOP RELAT R, P120
  • [3] BLUNN GW, 1991, CLIN ORTHOP RELAT R, P253
  • [4] COLLIER JP, 1991, CLIN ORTHOP RELAT R, P232
  • [5] MOTION OF THE KNEE AFTER CONDYLAR RESURFACING - AN INVIVO STUDY
    ELNAHASS, B
    MADSON, MM
    WALKER, PS
    [J]. JOURNAL OF BIOMECHANICS, 1991, 24 (12) : 1107 - 1117
  • [6] POLYETHYLENE WEAR OF METAL-BACKED TIBIAL COMPONENTS IN TOTAL AND UNICOMPARTMENTAL KNEE PROSTHESES
    ENGH, GA
    DWYER, KA
    HANES, CK
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (01): : 9 - 17
  • [7] INFRARED MICROSPECTROMETRIC STUDY OF THE THERMAL HISTORY OF FUSION-WELDING OF POLYETHYLENE
    GUEUGNAUT, D
    WACKERNIE, P
    FORGERIT, JP
    [J]. JOURNAL OF APPLIED POLYMER SCIENCE, 1988, 35 (06) : 1683 - 1694
  • [8] RETRIEVAL ANALYSIS OF TOTAL KNEE PROSTHESES - A METHOD AND ITS APPLICATION TO 48 TOTAL CONDYLAR PROSTHESES
    HOOD, RW
    WRIGHT, TM
    BURSTEIN, AH
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1983, 17 (05): : 829 - 842
  • [9] COMBINED CHEMICAL AND MECHANICAL EFFECTS ON FREE-RADICALS IN UHMWPE JOINTS DURING IMPLANTATION
    JAHAN, MS
    WANG, C
    SCHWARTZ, G
    DAVIDSON, JA
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1991, 25 (08): : 1005 - 1017
  • [10] POLYETHYLENE WEAR IN UNCEMENTED KNEE REPLACEMENTS
    JONES, SMG
    PINDER, IM
    MORAN, CG
    MALCOLM, AJ
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (01): : 18 - 22