Cooled-tip ablation results in increased radiofrequency power delivery and lesion size in the canine heart: Importance of catheter-tip temperature monitoring for prevention of popping and impedance rise

被引:53
作者
Watanabe, I [1 ]
Masaki, R [1 ]
Min, N [1 ]
Oshikawa, N [1 ]
Okubo, K [1 ]
Sugimura, H [1 ]
Kojima, T [1 ]
Saito, S [1 ]
Ozawa, Y [1 ]
Kanmatsuse, K [1 ]
机构
[1] Nihon Univ, Sch Med, Dept Med 2, Itabashi Ku, Tokyo 1738610, Japan
关键词
catheter ablation; cooled-tip electrode; lesion size; impedance rise; popping;
D O I
10.1023/A:1014140104777
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Since myocardial lesion size during radio-frequency (RF) ablation is limited at high power by impedance rise when electrode tip temperature exceed 100 degreesC, controlling tip temperature by continuous intraelectrode saline infusion could permit generation of larger lesion. (1) Two dogs randomly received either standard or cooled tip RF ablation at 4 to 6 separate LV sites. Power output of 30 W was delivered via modified 7 Fr deflectable catheter with 4 mm tip for up to 120 sec or until impedance rise occurred. (2) Six dogs randomly received cooled tip RF ablation at power output of 20, 30, 40 W for 120 sec. (3) Three dogs randomly received cooled tip RF ablation using room temperature saline (21-25 degreesC) or chilled saline (1-4 degreesC) infusion. Results: Overall, peak tip temperature was lower for cooled vs standard RF deliveries (97+/-17 degreesC vs. 42+/-8 degreesC). Lesion depth and volume were significantly larger for cooled burns. Lesion depth and volume and the incidence of abrupt impedance rise/popping did not differ between room temperature saline and chilled saline infusion. The catheter-tip temperature at the onset of popping and abrupt impedance rise was 54+/-5 degreesC(48-60 degreesC) and 59+/-10 degreesC(50-75 degreesC). Conclusion: Cooled tip RF current delivery at high power is associated with increased myocardial lesion size which may facilitate successful ablation of ventricular tachycardia associated with acquired structural heart disease. Catheter-tip temperature should be maintained below 45 degreesC to prevent popping and abrupt impedance rise during RF energy delivery.
引用
收藏
页码:9 / 16
页数:8
相关论文
共 30 条
[1]  
BERGAU D, 1993, CIRCULATION, V88, P64
[2]   Catheter ablation of ventricular tachycardia in patients with structural heart disease using cooled radiofrequency energy - Results of a prospective multicenter study [J].
Calkins, H ;
Epstein, A ;
Packer, D ;
Arria, AM ;
Hummel, J ;
Gilligan, DM ;
Trusso, J ;
Carlson, M ;
Luceri, R ;
Kopelman, H ;
Wilber, D ;
Wharton, JM ;
Stevenson, W .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2000, 35 (07) :1905-1914
[3]   Ablation of ventricular tachycardia with a saline-cooled radiofrequency catheter: Anatomic and histologic characteristics of the lesions in humans [J].
Delacretaz, E ;
Stevenson, WG ;
Winters, GL ;
Mitchell, RN ;
Stewart, S ;
Lynch, K ;
Friedman, PL .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1999, 10 (06) :860-865
[4]   ENDOCARDIAL MAPPING OF VENTRICULAR-TACHYCARDIA IN THE INTACT HUMAN HEART .2. EVIDENCE FOR MULTIUSE REENTRY IN A FUNCTIONAL SHEET OF SURVIVING MYOCARDIUM [J].
DOWNAR, E ;
KIMBER, S ;
HARRIS, L ;
MICKLEBOROUGH, L ;
SEVAPTSIDIS, E ;
MASSE, S ;
CHEN, TCK ;
GENGA, A .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 20 (04) :869-878
[5]  
FITZGERALD DM, 1991, J CARDIOVASC ELECTR, V2, P193
[6]   OBSERVATIONS ON ELECTRODE-TISSUE INTERFACE TEMPERATURE AND EFFECT ON ELECTRICAL-IMPEDANCE DURING RADIOFREQUENCY ABLATION OF VENTRICULAR MYOCARDIUM [J].
HAINES, DE ;
VEROW, AF .
CIRCULATION, 1990, 82 (03) :1034-1038
[7]  
Huang SKS, 1989, CIRCULATION, V80, pII
[8]   RADIOFREQUENCY CURRENT DIRECTED ACROSS THE MITRAL ANULUS WITH A BIPOLAR EPICARDIAL-ENDOCARDIAL CATHETER ELECTRODE CONFIGURATION IN DOGS [J].
JACKMAN, WM ;
KUCK, KH ;
NACCARELLI, GV ;
CARMEN, L ;
PITHA, J .
CIRCULATION, 1988, 78 (05) :1288-1298
[9]   Prospective randomized comparison of irrigated-tip versus conventional-tip catheters for ablation of common flutter [J].
Jaïs, P ;
Shah, DC ;
Haissaguerre, M ;
Hocini, M ;
Garrigue, S ;
Le Metayer, P ;
Clémenty, J .
CIRCULATION, 2000, 101 (07) :772-776
[10]  
JAIS P, SUCCESSFUL IRRIGATED