Transesophageal cardiac pacing during magnetic resonance imaging: Feasibility and safety considerations

被引:16
作者
Hofman, MBM
deCock, CC
vanderLinden, JC
vanRossum, AC
Visser, FC
Sprenger, M
Westerhof, N
机构
[1] FREE UNIV AMSTERDAM,ICAR VU,DEPT CLIN PHYS & ENGN,AMSTERDAM,NETHERLANDS
[2] FREE UNIV AMSTERDAM,ICAR VU,DEPT PHYSIOL,AMSTERDAM,NETHERLANDS
[3] FREE UNIV AMSTERDAM,ICAR VU,DEPT CARDIOL,AMSTERDAM,NETHERLANDS
[4] FREE UNIV AMSTERDAM,ICAR VU,DEPT PATHOL,AMSTERDAM,NETHERLANDS
关键词
cardiac pacing; safety; magnetic resonance imaging; heating effects;
D O I
10.1002/mrm.1910350320
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The feasibility and safety of transesophageal cardiac pacing during clinical MRI at 1.5 Tesla is considered, An MRI compatible pace catheter was developed. In vitro testing showed a normal performance of the pulse generator, image artifacts that extended less than 11 mm from the catheter, and a less than 5% increase in noise, Cardiac stimulation induced by MRI was not observed and, theoretically, is not expected, Potentially, tissue around the catheter tip may become heated, This heating (Delta T) was monitored, Eight dogs were exposed to MRI during pacing, For low RF radiation exposure, a time-averaged squared B-1 field below 0.08 pT(2) (SAR < 0.03 W/kg), Delta T was below 1 degrees C, For high RF radiation exposure, but at normal RF radiation specific absorption rate (0.4 W/kg), Delta T was 5 degrees C, Thus, transesophageal atrial pacing during MRI at low RF exposure seems to be possible to perform cardiac stress studies or to correct unstable heart rates.
引用
收藏
页码:413 / 422
页数:10
相关论文
共 52 条
[41]   CARDIAC PACING WITH AN ESOPHAGEAL ELECTRODE [J].
ROWE, GG ;
TERRY, W ;
NEBLETT, I .
AMERICAN JOURNAL OF CARDIOLOGY, 1969, 24 (04) :548-&
[42]   A NEW DEVICE FOR EXERCISE MR IMAGING [J].
SCHAEFER, S ;
PESHOCK, RM ;
PARKEY, RW ;
WILLERSON, JT .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1986, 147 (06) :1289-1290
[43]   ALTERATIONS IN BODY AND SKIN TEMPERATURES CAUSED BY MAGNETIC-RESONANCE IMAGING - IS THE RECOMMENDED EXPOSURE FOR RADIOFREQUENCY RADIATION TOO CONSERVATIVE [J].
SHELLOCK, FG ;
SCHAEFER, DJ ;
CRUES, JV .
BRITISH JOURNAL OF RADIOLOGY, 1989, 62 (742) :904-909
[44]   MR PROCEDURES AND BIOMEDICAL IMPLANTS, MATERIALS, AND DEVICES - 1993 UPDATE [J].
SHELLOCK, FG ;
MORISOLI, S ;
KANAL, E .
RADIOLOGY, 1993, 189 (02) :587-599
[45]   WIRELESS RETROSPECTIVE GATING - APPLICATION TO CINE CARDIAC IMAGING [J].
SPRAGGINS, TA .
MAGNETIC RESONANCE IMAGING, 1990, 8 (06) :675-681
[46]  
Stratton J. A., 1941, International series in physics
[47]  
TABER K, 1990, P SMRM 9 ANN M, P608
[48]  
*US FDA, 1982, GUID EV EL EXP TRIAL
[49]   REGIONAL MYOCARDIAL-METABOLISM OF HIGH-ENERGY PHOSPHATES DURING ISOMETRIC-EXERCISE IN PATIENTS WITH CORONARY-ARTERY DISEASE [J].
WEISS, RG ;
BOTTOMLEY, PA ;
HARDY, CJ ;
GERSTENBLITH, G .
NEW ENGLAND JOURNAL OF MEDICINE, 1990, 323 (23) :1593-1600
[50]   CONTRAST-ENHANCED 1ST PASS MYOCARDIAL PERFUSION IMAGING - CORRELATION BETWEEN MYOCARDIAL BLOOD-FLOW IN DOGS AT REST AND DURING HYPEREMIA [J].
WILKE, N ;
SIMM, C ;
ZHANG, J ;
ELLERMANN, J ;
YA, X ;
MERKLE, H ;
PATH, G ;
LUDEMANN, H ;
BACHE, RJ ;
UGURBIL, K .
MAGNETIC RESONANCE IN MEDICINE, 1993, 29 (04) :485-497