Mechanical assistance for enhancement of in vitro effectiveness for hydrodynamic thrombectomy

被引:7
作者
Müller-Hülsbeck, S
Grimm, J
Suwelack, D
Heller, M
机构
[1] Univ Hosp Kiel, Dept Radiol, D-24105 Kiel, Germany
[2] Univ Kiel, Dept Med Stat, Kiel, Germany
关键词
thrombectomy; rheolytic/hydrodynamic; thrombosis; experimental; interventional procedures; complications;
D O I
10.1097/00004424-199911000-00002
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. TO determine whether the efficacy for clot removal and procedure-related peripheral particle embolization for the hydrodynamic thrombectomy device CF 105 AngioJet (CF 105) can be enhanced with the use of guiding catheters and an occlusion balloon in an in vitro how model. METHODS. Thrombectomy of clots (n = 97) from 5-day-old human blood (range, 9.28-9.77 g) was performed with the CP 105 alone, with additional guide wires (CF 1050.014/0.018), with a 30 degrees angled tip (CF 10530 degrees), with guiding catheters (CF 105MP/HS), and with an additional occlusion balloon (CF 105OB) in a flow model (flow 1 L/min) of silicone tubes (inner tube diameter, 7 mm), RESULTS. Mean thrombectomy time ranged from 28.71 seconds (CF 105OB) to 37.02 seconds (CF 10530 degrees), When a guide wire was used, the time for thrombectomy was not prolonged. None of the devices tested worked isovolumetrically. The ratio of applied saline and aspirated fluid (0.34) for CF 105 alone was significantly better than for CF 1050.014/0.018 (0.49/0.60; for both P < 0.001). Compared with CF 105 alone (273.16 mg), wall-adhesive thrombus was reduced significantly with CF 105MP (43.66 mg), CF 105HS (48.66 mg), and CF 10530 degrees (97.27 mg; P < 0.001 for all). Compared with the highest rate of embolism for CF 105 alone (4.62%), CF 105OB (2.50%; P < 0.001), CF 105HS (3.04%; P < 0.001), CF 105MP (3.49%; P < 0.001), and CF 105018 (3.64%; P < 0.001) produced better ratios. CONCLUSIONS. The use of guiding catheters and occlusion balloons is feasible to enhance hydrodynamic thrombectomy in vitro. This combination reduces procedure-related peripheral particle embolization and wall-adhesive thrombus during hydrodynamic thrombectomy, To avoid blood loss, the device should be used only with a 0.018 " guide wire.
引用
收藏
页码:669 / 677
页数:9
相关论文
共 32 条
[31]  
Yasui K, 1993, J Vasc Interv Radiol, V4, P275, DOI 10.1016/S1051-0443(93)71853-9
[32]   PRESSURE DROPS ACROSS PROSTHETIC AORTIC HEART-VALVES UNDER STEADY AND PULSATILE FLOW - INVITRO MEASUREMENTS [J].
YOGANATHAN, AP ;
CORCORAN, WH ;
HARRISON, EC .
JOURNAL OF BIOMECHANICS, 1979, 12 (02) :153-164