Mixed-oxides prosthetic ceramic ball heads.: Part 2:: Effect of the ZrO2 fraction on the wear of ceramic on ceramic joints

被引:41
作者
Affatato, S
Testoni, M
Cacciari, GL
Toni, A
机构
[1] Ist Ortoped Rizzoli, Lab Tecnol Med, I-40136 Bologna, Italy
[2] Off Ortoped Rizzoli, Bologna, Italy
[3] Univ Bologna, Clin Ortoped, Bologna, Italy
关键词
hip joint simulator; lubricant; pure alumina; zirconia; wear;
D O I
10.1016/S0142-9612(99)00093-9
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Three different types of mixed-oxides ceramic ball heads have been investigated for their wear behaviour against acetabular cups of the same materials in a hip joint simulator. Mixed-oxides ceramics have been indicated in literature as a promising compromise between strength and wear but no reports are available on the influence of a percentage of zirconia in a ceramic femoral head when sliding against itself. Mixed-oxides ceramic acetabular cups and femoral heads were tested on a simulator apparatus with a sinusoidal load in presence of bovine calf serum. The experimental results did not show any significant difference between the experimental and commercial ceramic material couplings. These results were found to be in accord with those developed in Part 1. (C) 1999 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:1925 / 1929
页数:5
相关论文
共 16 条
[1]   Wear behaviour of alumina toughened zirconia materials [J].
Cherif, K ;
Gueroult, B ;
Rigaud, M .
WEAR, 1996, 199 (01) :113-121
[2]  
CUCKLER JM, 1995, CLIN ORTHOP RELAT R, P57
[3]  
DEBORSHIRE B, 1994, MED ENG PHYS, V16, P229
[4]   Tribological investigations of the wear couple alumina-CFRP for total hip replacement [J].
Früh, HJ ;
Willmann, G .
BIOMATERIALS, 1998, 19 (13) :1145-1150
[5]  
GUEROULT B, 1994, CAN CERAM, V63, P132
[6]   PARTICULATE DEBRIS AND FAILURE OF TOTAL HIP REPLACEMENTS [J].
JASTY, M .
JOURNAL OF APPLIED BIOMATERIALS, 1993, 4 (03) :273-276
[7]   LOW WEAR RATE OF UHMWPE AGAINST ZIRCONIA CERAMIC (Y-PSZ) IN COMPARISON TO ALUMINA CERAMIC AND SUS 316L ALLOY [J].
KUMAR, P ;
OKA, M ;
IKEUCHI, K ;
SHIMIZU, K ;
YAMAMURO, T ;
OKUMURA, H ;
KOTOURA, Y .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1991, 25 (07) :813-828
[8]  
MCKELLOP H, 1992, 4TH WORLD BIOM C BER, P118
[9]   ON THE TRUE WEAR RATE OF ULTRA HIGH-MOLECULAR-WEIGHT POLYETHYLENE IN THE TOTAL HIP-PROSTHESIS [J].
ROSE, RM ;
NUSBAUM, HJ ;
SCHNEIDER, H ;
RIES, M ;
PAUL, I ;
CRUGNOLA, A ;
SIMON, SR ;
RADIN, EL .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1980, 62 (04) :537-549
[10]   WEAR OF THE POLYETHYLENE ACETABULAR CUP - METALLIC AND CERAMIC HEADS COMPARED IN A HIP SIMULATOR [J].
SAIKKO, VO ;
PAAVOLAINEN, PO ;
SLATIS, P .
ACTA ORTHOPAEDICA SCANDINAVICA, 1993, 64 (04) :391-402