A simple cassette as point-of-care diagnostic device for naked-eye colorimetric bacteria detection

被引:87
作者
Safavieh, Mohammadali [1 ]
Ahmed, Minhaz Uddin [1 ,2 ]
Sokullu, Esen [1 ]
Ng, Andy [1 ]
Braescuac, Liliana [1 ,3 ]
Zourob, Mohammed [4 ]
机构
[1] Univ Quebec, INRS EMT, Varennes, PQ J3X 1S2, Canada
[2] Univ Brunei Darussalam, Fac Sci, BE-1410 Gadong, Brunei
[3] West Univ Timisoara, Dept Comp Sci, Timisoara 300223, Romania
[4] Cranfield Univ, Ctr Biomed Engn, Cranfield MK43 0AL, Beds, England
基金
加拿大自然科学与工程研究理事会;
关键词
MEDIATED ISOTHERMAL AMPLIFICATION; RAPID DETECTION; TECHNOLOGIES; PCR; ASSAY; LAMP;
D O I
10.1039/c3an01859h
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Effective pathogen detection is necessary for treatment of infectious diseases. Point of care (POC) devices have tremendously improved the global human heath. However, design criteria for sample processing POC devices for pathogen detection in limited infrastructure are challenging and can make a significant contribution to global health by providing rapid and sensitive detection of bacteria in food, water, and patient samples. In this paper, we demonstrate a novel portable POC diagnostic device that is simple to assemble for genetic detection of bacterial pathogens by isothermal DNA amplification. The device is fabricated with very low production cost, using simple methods and easy-to-access materials on a flexible ribbon polyethylene substrate. We showed that the device is capable of detection of 30 CFU mL(-1) of E. coli and 200 CFU mL(-1) of S. aureus in less than 1 hour. Through numerical simulations, we estimated that the device can be extended to high-throughput detection simultaneously performing a minimum of 36 analyses. This robust and sensitive detection device can be assembled and operated by non-specialist personnel, particularly for multiple bacterial pathogen detections in low-resource settings.
引用
收藏
页码:482 / 487
页数:6
相关论文
共 40 条
[1]   Electrochemical biosensors for medical and food applications [J].
Ahmed, Minhaz Uddin ;
Hossain, Mohammad Mosharraf ;
Tamiya, Eiichi .
ELECTROANALYSIS, 2008, 20 (06) :616-626
[2]   Real-time electrochemical detection of pathogen DNA using electrostatic interaction of a redox probe [J].
Ahmed, Minhaz Uddin ;
Nahar, Sharifun ;
Safavieh, Mohammadali ;
Zourob, Mohammed .
ANALYST, 2013, 138 (03) :907-915
[3]   Helicase-dependent amplification: use in OnChip amplification and potential for point-of-care diagnostics [J].
Andresen, Dennie ;
von Nickisch-Rosenegk, Markus ;
Bier, Frank F. .
EXPERT REVIEW OF MOLECULAR DIAGNOSTICS, 2009, 9 (07) :645-650
[4]  
[Anonymous], CAUS DEATH 2008 DAT
[5]  
Craw P, 2012, LAB CHIP, V12, P2469, DOI [10.1039/c21c40100b, 10.1039/c2lc40100b]
[6]   Continuous-flow thermal gradient PCR [J].
Crews, Niel ;
Wittwer, Carl ;
Gale, Bruce .
BIOMEDICAL MICRODEVICES, 2008, 10 (02) :187-195
[7]   Isothermal Amplification Using a Chemical Heating Device for Point-of-Care Detection of HIV-1 [J].
Curtis, Kelly A. ;
Rudolph, Donna L. ;
Nejad, Irene ;
Singleton, Jered ;
Beddoe, Andy ;
Weigl, Bernhard ;
LaBarre, Paul ;
Owen, S. Michele .
PLOS ONE, 2012, 7 (02)
[8]  
Eiwary S., 2008, PRINCIPLES BACTERIAL
[9]   Colorimetric detection of loop-mediated isothermal amplification reaction by using hydroxy naphthol blue [J].
Goto, Motoki ;
Honda, Eiichi ;
Ogura, Atsuo ;
Nomoto, Akio ;
Hanaki, Ken-Ichi .
BIOTECHNIQUES, 2009, 46 (03) :167-+
[10]   Point of Care Diagnostics: Status and Future [J].
Gubala, Vladimir ;
Harris, Leanne F. ;
Ricco, Antonio J. ;
Tan, Ming X. ;
Williams, David E. .
ANALYTICAL CHEMISTRY, 2012, 84 (02) :487-515