Importance of injection solution composition for LC-MS-MS methods

被引:43
作者
Weng, ND [1 ]
Chen, YL [1 ]
Shou, W [1 ]
Jiang, XY [1 ]
机构
[1] Covance Labs Inc, Dept Bioanalyt Chem, Madison, WI 53707 USA
关键词
injection solution; LC-MS-MS;
D O I
10.1016/S0731-7085(01)00439-3
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
For the first time, the influence of the injection solution composition on the quality of LC-MS-MS methods, in terms of column efficiency and peak shape, was systematically investigated. Various types of compounds, including polar ionic acidic, polar ionic basic and non-polar neutral compounds, were prepared in different solutions ranging from 100% water to 100% acetonitrile, Different volumes of these solutions were injected onto either C-18 or silica columns connected to tandem mass spectrometry. The mobile phases consisted of acetonitrile, water, and small amounts of volatile acid or buffer. On silica columns, the influence of injection solution on the peak shape and column efficiency was straightforward. The sharpest peaks and the highest column efficiency were obtained with 100% acetonitrile as the injection solvent. On C-18 columns, this type of influence was less clear due to the dual retention mechanism of the bonded phase and of the residual silanol groups. On C-18 column, retention due to residual silanol groups was significant even with a mobile phase containing less than 50%, acetonitrile. Poor peak shape was observed when the injection solution had a stronger eluting strength than mobile phase, particularly for early eluting peaks, (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:753 / 767
页数:15
相关论文
共 22 条
[1]  
[Anonymous], 1997, ELECTROSPRAY IONIZAT
[2]   Automated protein precipitation by filtration in the 96-well format [J].
Biddlecombe, RA ;
Pleasance, S .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 1999, 734 (02) :257-265
[3]   Novel high-performance liquid chromatographic and solid-phase extraction methods for quantitating methadone and its metabolite in spiked human urine [J].
Cheng, YF ;
Neue, UD ;
Woods, LL .
JOURNAL OF CHROMATOGRAPHY B, 1999, 729 (1-2) :19-31
[4]   Detection and confirmation of beta-agonists in bovine retina using LC/APCI-MS [J].
Doerge, DR ;
Churchwell, MI ;
Holder, CL ;
Rowe, L ;
Bajic, S .
ANALYTICAL CHEMISTRY, 1996, 68 (11) :1918-1923
[5]   EQUATIONS FOR CALCULATION OF CHROMATOGRAPHIC FIGURES OF MERIT FOR IDEAL AND SKEWED PEAKS [J].
FOLEY, JP ;
DORSEY, JG .
ANALYTICAL CHEMISTRY, 1983, 55 (04) :730-737
[6]   Effect of the sample matrix on the determination of indinavir in human urine by HPLC with turbo ion spray tandem mass spectrometric detection [J].
Fu, I ;
Woolf, EJ ;
Matuszewski, BK .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1998, 18 (03) :347-357
[7]  
Joyce KB, 1998, RAPID COMMUN MASS SP, V12, P1899, DOI 10.1002/(SICI)1097-0231(19981215)12:23<1899::AID-RCM417>3.0.CO
[8]  
2-I
[9]  
Lee MS, 1999, MASS SPECTROM REV, V18, P187, DOI 10.1002/(SICI)1098-2787(1999)18:3/4<187::AID-MAS2>3.0.CO
[10]  
2-K