Inherent variability of CT lung nodule measurements in vivo using semiautomated volumetric measurements
被引:126
作者:
Goodman, LR
Gulsun, M
Washington, L
Nagy, PG
Piacsek, KL
机构:
[1] Med Coll Wisconsin, Dept Radiol, Milwaukee, WI 53226 USA
[2] GE Healthcare, Waukesha, WI 53188 USA
关键词:
chest imaging;
CT;
CT image processing;
CT technique;
lung;
lung cancer;
lung nodule;
D O I:
10.2214/AJR.04.1821
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
OBJECTIVE. The objective of our study was to evaluate repeatability and reproducibility of lung nodule volume measurements using volumetric nodule-sizing software. MATERIALS AND METHODS. Fifty nodules, less than 20 mm in diameter, in 29 patients were scanned with 1.25-mm collimation using MDCT (time I = T1). During the same session, two additional scans, using identical technique, were obtained through each nodule (T2, T3). Three observers working independently then obtained volumetric measurements using a semiautomated volumetric nodule-sizing software package. Qualitative nodule characterization was also performed. The Bland-Altman method for assessing measurement agreement was used to calculate the 95% limits for agreement for nodule volumes at T1, T2, and T3. RESULTS. Automated nodule segmentation was successful in 438 (97%) of 450 measurements. Forty-three nodules were available for final evaluation. Twenty-six nodules had well-defined edges, and 17 had irregular or spiculated margins. Seventeen were freestanding, 16 were juxtapleural, and 10 were juxtavascular in location. Average nodule volume was 345.5 mm(3) (range, 49.3-1,434 mm(3)). The mean interobserver variability (repeatability) was 0.018% (SD = 0.73%), and the SD of the mean for the three contemporaneous scans (reproducibility) was 13.1% (confidence limits, +/- 25.6%). SD and confidence limits narrowed as volumes increased. CONCLUSION. Volumetric measurements show minimal interobserver variability (0.018%) but an interscan SEM of 13.1% (confidence limits, +/- 25.6%). Repeatability and reproducibility of volumetric measurements are better than those of linear measurements reported in the literature.
引用
收藏
页码:989 / 994
页数:6
相关论文