Compression depth estimation for CPR quality assessment using DSP on accelerometer signals

被引:105
作者
Aase, SO
Myklebust, H
机构
[1] Stavanger Univ Coll, Dept Elect & Comp Engn, N-4004 Stavanger, Norway
[2] Laerdal Med, N-4002 Stavanger, Norway
关键词
accelerometer; compression depth; resuscitation;
D O I
10.1109/10.983461
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Chest compression is a vital part of cardiopulmonary resuscitation (CPR). This paper demonstrates how the compression depth can be estimated using the principles of inertia navigation. The proposed method uses accelerometer sensors, one placed on the patient's chest, the other beside the patient. The acceleration-to-position conversion is performed using discrete-time digital signal processing (DSP). Instability problems due to integration are combated using a set of boundary conditions. The proposed algorithm is tested on a mannequin in harsh environments, where the patient Is exposed to external forces as in a boat or car, as well as improper sensor/patient alignment. The overall performance is an estimation depth error of 4.3 mm in these environments, which is reduced to 1.6 mm in a regular, flat-floor controlled environment.
引用
收藏
页码:263 / 268
页数:6
相关论文
共 13 条
[11]   Cardiopulmonary resuscitation: errors made by pre-hospital emergency medical personnel [J].
Liberman, M ;
Lavoie, A ;
Mulder, D ;
Sampalis, J .
RESUSCITATION, 1999, 42 (01) :47-55
[12]  
Proakis J. G, 1996, Digital Signal Processing: Principles, Algorithms, and Applications, V3rd
[13]   QUALITY OF BYSTANDER CARDIOPULMONARY-RESUSCITATION INFLUENCES OUTCOME AFTER PREHOSPITAL CARDIAC-ARREST [J].
WIK, L ;
STEEN, PA ;
BIRCHER, NG .
RESUSCITATION, 1994, 28 (03) :195-203