Determination of water absorbed dose in a carbon ion beam using thimble ionization chambers

被引:73
作者
Hartmann, GH
Jäkel, O
Heeg, P
Karger, CP
Kriessbach, A
机构
[1] German Canc Res Ctr, Dept Med Phys, D-69009 Heidelberg, Germany
[2] Univ Heidelberg, Radiol Hosp, D-69120 Heidelberg, Germany
关键词
D O I
10.1088/0031-9155/44/5/008
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The method to measure absorbed dose to water in a field of carbon ions as applied for the heavy ion therapy project at the Heavy Ion Research Laboratory in Darmstadt (GSI), Germany, is described in detail. Thimble ionization chambers with a water absorbed calibration factor are applied. The dose obtained with this method was compared with that obtained at the heavy ion therapy facility HIMAC at the National Institute of Radiological Sciences in Chiba, Japan, using the Japanese code of practice. The agreement found was better than 1%. The combined uncertainty of the determination of absorbed dose to water was estimated to amount to 5%.
引用
收藏
页码:1193 / 1206
页数:14
相关论文
共 45 条
[31]   Monte Carlo study of fluence perturbation effects on cavity dose response in clinical proton beams [J].
Palmans, H ;
Verhaegen, F .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (01) :65-89
[32]   A MEASUREMENT OF W FOR 150 MEV PROTONS IN NITROGEN AND ARGON [J].
PETTI, PL ;
VERHEY, L ;
WILSON, R .
PHYSICS IN MEDICINE AND BIOLOGY, 1986, 31 (10) :1129-1138
[33]  
Scharf WH, 1996, PHYS MEDICA, V12, P199
[34]  
SCHIMMERLING W, 1982, P 8 S JUL GERM 27 SE, P311
[35]  
SCHWAB T, 1990, THESIS J LIEBIG U GI
[36]   DEDUCTION OF THE AIR W-VALUE IN A THERAPEUTIC PROTON-BEAM [J].
SIEBERS, JV ;
VATNITSKY, SM ;
MILLER, DW ;
MOYERS, MF .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (08) :1339-1356
[37]   MEASUREMENT OF AVERAGE ENERGY REQUIRED TO CREATE AN ION-PAIR IN NITROGEN BY 250 MEV-AMU C-6+ IONS [J].
STEPHENS, LD ;
THOMAS, RH ;
KELLY, LS .
PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (04) :570-576
[38]   W-VALUE MEASUREMENTS FOR PROTONS IN TISSUE-EQUIVALENT GAS AND ITS CONSTITUENT GASES [J].
THOMAS, DJ ;
BURKE, M .
PHYSICS IN MEDICINE AND BIOLOGY, 1985, 30 (11) :1201-1213
[39]   A MEASUREMENT OF THE AVERAGE ENERGY REQUIRED TO CREATE AN ION-PAIR IN NITROGEN BY HIGH-ENERGY IONS [J].
THOMAS, RH ;
LYMAN, JT ;
DECASTRO, TM .
RADIATION RESEARCH, 1980, 82 (01) :1-12
[40]  
TSUJII T, 1994, 20 PTCOG M CHEST CAM, P22