A sealless centrifugal blood pump with passive magnetic and hydrodynamic bearings

被引:23
作者
Wampler, R [1 ]
Lancisi, D [1 ]
Indravudh, V [1 ]
Gauthier, R [1 ]
Fine, R [1 ]
机构
[1] Kriton Med Inc, Citrus Hts, CA 95610 USA
关键词
magnetically suspended centrifugal pump; ventricular assist device; axial flux gap motor; radial magnetic bearing; hydrodynamic bearings;
D O I
10.1046/j.1525-1594.1999.06422.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We are developing a permanently implantable ventricular assist system based on a sealless centrifugal blood pump. The impeller of the pump is supported by a passive radial magnetic bearing acting in synergy with hydrodynamic bearings. Torque is transmitted to the impeller by electromagnetic coupling via an integrated axial flux gap motor. Computer modeling has been used extensively to guide the hydraulic and electromagnetic design of the pump. As part of the development effort, a prototype system was built, which consisted of a radial magnetic bearing, an axial air gap motor, and a pivot bearing to constrain the axial motion. The following testing has been completed to validate the design. First, hydraulic tests have demonstrated sufficient hydraulic performance. Second, preliminary in vitro evaluation of hemolysis was low compared to that of a BioPump control. Third, a 6 h in vivo experiment was successfully completed.
引用
收藏
页码:780 / 784
页数:5
相关论文
共 17 条
[1]   CENTRIFUGAL BLOOD PUMP WITH A MAGNETICALLY SUSPENDED IMPELLER [J].
AKAMATSU, T ;
NAKAZEKI, T ;
ITOH, H .
ARTIFICIAL ORGANS, 1992, 16 (03) :305-308
[2]  
BEARSON GB, 1995, ASAIO J, V41, P49
[3]   24-HOUR LEFT-VENTRICULAR BYPASS WITH A CENTRIFUGAL BLOOD PUMP [J].
BERNSTEIN, EF ;
DELARIA, GA ;
JOHANSEN, KH ;
SHUMAN, RL ;
STASZ, P ;
REICH, S .
ANNALS OF SURGERY, 1975, 181 (04) :412-417
[4]   A FREE-FLOATING BODY AS THE ROTOR OF A CENTRIFUGAL BLOOD PUMP FOR LVAD OR TAH [J].
BRAMM, G ;
NOVAK, P ;
OLSEN, DB .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1982, 5 (03) :161-162
[5]   Development of the Nimbus/Pittsburgh axial flow left ventricular assist system [J].
Butler, K ;
Thomas, D ;
Antaki, J ;
Borovetz, H ;
Griffith, B ;
Kameneva, M ;
Kormos, R ;
Litwak, P .
ARTIFICIAL ORGANS, 1997, 21 (07) :602-610
[6]  
DAMM G, 1994, ARTIF ORGANS, V118, P44
[7]  
DORMAN F, 1969, T AM SOC ART INT ORG, V15, P441
[8]   The Cleveland clinic rotodynamic pump program [J].
Golding, LAR ;
Smith, WA ;
Bodmann, DR .
ARTIFICIAL ORGANS, 1996, 20 (06) :481-484
[9]  
GOLDING LR, 1978, T AM SOC ART INT ORG, V24, P93
[10]  
HART R, 1995, T AM SOC ART INT ORG, V41, P42