An in vitro analysis of the output of salbutamol from different nebulizers

被引:29
作者
Barry, PW [1 ]
O'Callaghan, C [1 ]
机构
[1] Univ Leicester, Dept Child Hlth, Leicester LE1 7RH, Leics, England
关键词
drug administration - inhalation; nebulizers; salbutamol;
D O I
10.1034/j.1399-3003.1999.13e37.x
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
The objective of this study was to determine the particle size and mass output of salbutamol from different nebulizers used under simulated breathing conditions. Seven nebulizer/compressor combinations were assessed. Each nebulizer was charged with 5 mg salbutamol solution and connected to a breathing simulator operating at tidal volumes of 150 mt and 600 mt. Nebulizers were operated for 15 min. Salbutamol collected on the filters was measured by liquid chromatography. Aerosol particle size was determined separately by laser diffraction. The Pari LC Star nebulizer delivered the most salbutamol at both tidal volumes. The maximal output of the Medicaid Ventstream and Sidestream nebulizers was two-thirds that of the LC Star, and they delivered less salbutamol than the LC Star or LC Plus nebulizers. The Intersurgical Cirrus nebulizer delivered the least salbutamol at both tidal volumes, although there was only a small difference between the Cirrus and Ventstream or Sidestream nebulizers at 150 mt tidal volume. The LC Plus nebulizer produced larger particles, mass median diameter 5.3 mu m, compared with 3.6-4.0 mu m for the other nebulizers. In conclusion, there were large differences in the delivery of salbutamol between the nebulizers studied, even between nebulizers of apparently the same class, and this should be borne in mind by regulatory authorities, clinicians and researchers.
引用
收藏
页码:1164 / 1169
页数:6
相关论文
共 27 条
[1]   PULMONARY DEPOSITION OF AEROSOLS IN CHILDREN WITH CYSTIC-FIBROSIS [J].
ALDERSON, PO ;
SECKERWA.RH ;
STROMINGER, DB ;
MARKHAM, J ;
HILL, RL .
JOURNAL OF PEDIATRICS, 1974, 84 (04) :479-484
[2]  
Bisgaard H, 1997, EUR RESPIR REV, V7, P376
[3]  
*BRIT STAND I, 1994, BS7711 BRIT STAND 3
[4]  
BUMILLER MR, 1997, INT SOC AER MED C P
[5]  
CARLSEN KCL, 1992, ARCH DIS CHILD, V67, P1077
[6]  
CHUNG KF, 1988, EUR RESPIR J, V1, P890
[7]   JET AND ULTRASONIC NEBULIZER OUTPUT - USE OF A NEW METHOD FOR DIRECT MEASUREMENT OF AEROSOL OUTPUT [J].
DENNIS, JH ;
STENTON, SC ;
BEACH, JR ;
AVERY, AJ ;
WALTERS, EH ;
HENDRICK, DJ .
THORAX, 1990, 45 (10) :728-732
[8]  
Denyer J, 1993, EUR RESP J S17, V6, p148S
[9]   Comparison of drug delivery from conventional versus "Venturi" nebulizers [J].
Devadason, SG ;
Everard, ML ;
Linto, JM ;
Le Souef, PN .
EUROPEAN RESPIRATORY JOURNAL, 1997, 10 (11) :2479-2483
[10]   STUDIES USING RADIOLABELED AEROSOLS IN CHILDREN [J].
EVERARD, ML .
THORAX, 1994, 49 (12) :1259-1266