The Monte Carlo-based inverse model of diffuse reflectance described in part I of this pair of companion papers was applied to the diffuse reflectance spectra of a set of 17 malignant and 24 normal-benign ex vivo human breast tissue samples. This model allows extraction of physically meaningful tissue parameters, which include the concentration of absorbers and the size and density of scatterers present in tissue. It was assumed that intrinsic absorption could be attributed to oxygenated and deoxygenated hemoglobin and beta-carotene, that scattering could be modeled by spheres of a uniform size distribution, and that the refractive indices of the spheres and the surrounding medium are known. The tissue diffuse reflectance spectra were evaluated over a wavelength range of 400-600 nm. The extracted parameters that showed the statistically most significant differences between malignant and nonmalignant breast tissues were hemoglobin saturation and the mean reduced scattering coefficient. Malignant tissues showed decreased hemoglobin saturation and an increased mean reduced scattering coefficient compared with nonmalignant tissues. A support vector machine classification algorithm was then used to classify a sample as malignant or nonmalignant based on these two extracted parameters and produced a cross-validated sensitivity and specificity of 82% and 92%, respectively. (c) 2006 Optical Society of America.
引用
收藏
页码:1072 / 1078
页数:7
相关论文
共 29 条
[11]   Concentration and oxygen saturation of haemoglobin of 50 breast tumours determined by time-domain optical mammography [J].
Grosenick, D ;
Wabnitz, H ;
Moesta, KT ;
Mucke, J ;
Möller, M ;
Stroszczynski, C ;
Stössel, J ;
Wassermann, B ;
Schlag, PM ;
Rinneberg, H .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (07) :1165-1181
[12]   Time-domain optical mammography: initial clinical results on detection and characterization of breast tumors [J].
Grosenick, D ;
Moesta, KT ;
Wabnitz, H ;
Mucke, J ;
Stroszczynski, C ;
Macdonald, R ;
Schlag, PM ;
Rinneberg, H .
APPLIED OPTICS, 2003, 42 (16) :3170-3186
[13]  
Gunn S. R., 1998, SUPPORT VECTOR MACHI
[14]   Stereotactic, automated, large-core needle biopsy of nonpalpable breast lesions: False-negative and histologic underestimation rates after long-term follow-up [J].
Jackman, RJ ;
Nowels, KW ;
Rodriguez-Soto, J ;
Marzoni, FA ;
Finkelstein, SI ;
Shepard, MJ .
RADIOLOGY, 1999, 210 (03) :799-805
[15]  
Johnson R.A., 1988, Applied multivariate statistical analysis
[16]   Large-needle core biopsy: Nonmalignant breast abnormalities evaluated with surgical excision or repeat core biopsy [J].
Meyer, JE ;
Smith, DN ;
Lester, SC ;
DiPiro, PJ ;
Denison, CM ;
Harvey, SC ;
Christian, RL ;
Richardson, A ;
Ko, WD .
RADIOLOGY, 1998, 206 (03) :717-720
[17]   Mechanisms of light scattering from biological cells relevant to noninvasive optical-tissue diagnostics [J].
Mourant, JR ;
Freyer, JP ;
Hielscher, AH ;
Eick, AA ;
Shen, D ;
Johnson, TM .
APPLIED OPTICS, 1998, 37 (16) :3586-3593
[18]   Characterizing mammalian cells and cell phantoms by polarized backscattering fiber-optic measurements [J].
Mourant, JR ;
Johnson, TM ;
Freyer, JP .
APPLIED OPTICS, 2001, 40 (28) :5114-5123
[19]   Monte Carlo-based inverse model for calculating tissue optical properties. Part I: Theory and validation on synthetic phantoms [J].
Palmer, GM ;
Ramanujam, N .
APPLIED OPTICS, 2006, 45 (05) :1062-1071
[20]   Comparison of multiexcitation fluorescence and diffuse reflectance spectroscopy for the diagnosis of breast cancer (March 2003) [J].
Palmer, GM ;
Zhu, CF ;
Breslin, TM ;
Xu, FS ;
Gilchrist, KW ;
Ramanujam, N .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2003, 50 (11) :1233-1242