Automated Estimation of Abdominal Effective Diameter for Body Size Normalization of CT Dose

被引:20
作者
Cheng, Phillip M. [1 ]
机构
[1] USC, Keck Sch Med, Dept Radiol, Los Angeles, CA 90033 USA
关键词
Computed tomography; Radiation dose; Body imaging; Quality control; Image analysis; INFORMATION; EXTRACTION; IMAGES; INDEX;
D O I
10.1007/s10278-012-9525-z
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
Most CT dose data aggregation methods do not currently adjust dose values for patient size. This work proposes a simple heuristic for reliably computing an effective diameter of a patient from an abdominal CT image. Evaluation of this method on 106 patients scanned on Philips Brilliance 64 and Brilliance Big Bore scanners demonstrates close correspondence between computed and manually measured patient effective diameters, with a mean absolute error of 1.0 cm (error range +2.2 to -0.4 cm). This level of correspondence was also demonstrated for 60 patients on Siemens, General Electric, and Toshiba scanners. A calculated effective diameter in the middle slice of an abdominal CT study was found to be a close approximation of the mean calculated effective diameter for the study, with a mean absolute error of approximately 1.0 cm (error range +3.5 to -2.2 cm). Furthermore, the mean absolute error for an adjusted mean volume computed tomography dose index (CTDIvol) using a mid-study calculated effective diameter, versus a mean per-slice adjusted CTDIvol based on the calculated effective diameter of each slice, was 0.59mGy (error range 1.64 to -3.12 mGy). These results are used to calculate approximate normalized dose length product values in an abdominal CT dose database of 12,506 studies.
引用
收藏
页码:406 / 411
页数:6
相关论文
共 9 条
[1]
American Association of Physicist in Medicine, 2011, 204 AAPM TASK GROUP
[2]
Cheng PM, 2011, AM ROENTG RAY SOC AN
[3]
Automated Extraction of Radiation Dose Information for CT Examinations [J].
Cook, Tessa S. ;
Zimmerman, Stefan ;
Maidment, Andrew D. A. ;
Kim, Woojin ;
Boonn, William W. .
JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2010, 7 (11) :871-877
[4]
Patient Size Measured on CT Images as a Function of Age at a Tertiary Care Children's Hospital [J].
Kleinman, Patricia L. ;
Strauss, Keith J. ;
Zurakowski, David ;
Buckley, Kevin S. ;
Taylor, George A. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 194 (06) :1611-1619
[5]
Automated Extraction of Radiation Dose Information From CT Dose Report Images [J].
Li, Xinhua ;
Zhang, Da ;
Liu, Bob .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 196 (06) :W781-W783
[6]
CT Dose Index and Patient Dose: They Are Not the Same Thing [J].
McCollough, Cynthia H. ;
Leng, Shuai ;
Yu, Lifeng ;
Cody, Dianna D. ;
Boone, John M. ;
McNitt-Gray, Michael F. .
RADIOLOGY, 2011, 259 (02) :311-316
[7]
Comparison of different body size parameters for individual dose adaptation in body CT of adults [J].
Menke, J .
RADIOLOGY, 2005, 236 (02) :565-571
[8]
ACR Dose Index Registry [J].
Morin, Richard L. ;
Coombs, Laura P. ;
Chatfield, Mythreyi Bhargavan .
JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2011, 8 (04) :288-291
[9]
AP diameter shows the strongest correlation with CTDI and DLP in abdominal and chest CT [J].
Zarb, Francis ;
Rainford, Louise ;
McEntee, Mark F. .
RADIATION PROTECTION DOSIMETRY, 2010, 140 (03) :266-273