Computer tomography assessment of pedicle screw placement in thoracic spine: comparison between free hand and a generic 3D-based navigation techniques

被引:107
作者
Allam, Yasser [1 ]
Silbermann, J. [2 ]
Riese, F. [3 ]
Greiner-Perth, R. [4 ]
机构
[1] El Hadra Univ Hosp, Spine Unit, Alexandria, Egypt
[2] Waldklinikum, Dept Spine Surg, Gera, Germany
[3] Univ Heart Ctr, Dept Radiol, Leipzig, Germany
[4] Hochfranken, Dept Orthopaed & Neurosurg, Hof, Germany
关键词
Pedicle screws; Thoracic spine Image-guided surgery; Spine navigation; IMAGE GUIDANCE; ACCURACY; INSERTION; FIXATION; SYSTEM;
D O I
10.1007/s00586-012-2505-7
中图分类号
R74 [神经病学与精神病学];
学科分类号
100204 [神经病学];
摘要
Although pedicle screw fixation is a well-established technique for the lumbar spine, screw placement in the thoracic spine is more challenging because of the smaller pedicle size and more complex 3D anatomy. The intraoperative use of image guidance devices may allow surgeons a safer, more accurate method for placing thoracic pedicle screws while limiting radiation exposure. This generic 3D imaging technique is a new generation intraoperative CT imaging system designed without compromise to address the needs of a modern OR. The aim of our study was to check the accuracy of this generic 3D navigated pedicle screw implants in comparison to free hand technique described by Roy-Camille at the thoracic spine using CT scans. The material of this study was divided into two groups: free hand group (group I) (18 patients; 108 screws) and 3D group (27 patients; 100 screws). The patients were operated upon from January 2009 to March 2010. Screw implantation was performed during internal fixation for fractures, tumors, and spondylodiscitis of the thoracic spine as well as for degenerative lumbar scoliosis. The accuracy rate in our work was 89.8 % in the free hand group compared to 98 % in the generic 3D navigated group. In conclusion, 3D navigation-assisted pedicle screw placement is superior to free hand technique in the thoracic spine.
引用
收藏
页码:648 / 653
页数:6
相关论文
共 26 条
[1]
Computer-assisted fluoroscopic targeting system for pedicle screw insertion [J].
Choi, WW ;
Green, BA ;
Levi, ADO .
NEUROSURGERY, 2000, 47 (04) :872-878
[2]
Pedicle instrumentation in the thoracic spine - A morphometric and cadaveric study for placement of screws [J].
Cinotti, G ;
Gumina, S ;
Ripani, M ;
Postacchini, F .
SPINE, 1999, 24 (02) :114-119
[3]
Pedicle screw insertion: computed tomography versus fluoroscopic image guidance [J].
Fu, Tsai-Sheng ;
Wong, Chak-Bor ;
Tsai, Tsung-Ting ;
Liang, Yen-Chiu ;
Chen, Lih-Huei ;
Chen, Wen-Jer .
INTERNATIONAL ORTHOPAEDICS, 2008, 32 (04) :517-521
[4]
Intraoperative spinal navigation [J].
Holly, LT ;
Foley, KT .
SPINE, 2003, 28 (15) :S54-S61
[5]
Free hand pedicle screw placement in the thoracic spine: Is it safe? [J].
Kim, YJ ;
Lenke, LG ;
Bridwell, KH ;
Cho, YSS ;
Riew, KD .
SPINE, 2004, 29 (03) :333-342
[6]
Kim YJJ, 2005, NEUROL INDIA, V53, P512
[7]
Pedicle screw placement accuracy - A meta-analysis [J].
Kosmopoulos, Victor ;
Schizas, Constantin .
SPINE, 2007, 32 (03) :E111-E120
[8]
MORPHOMETRY OF THE THORACIC AND LUMBAR SPINE RELATED TO TRANSPEDICULAR SCREW PLACEMENT FOR SURGICAL SPINAL FIXATION [J].
KRAG, MH ;
WEAVER, DL ;
BEYNNON, BD ;
HAUGH, LD .
SPINE, 1988, 13 (01) :27-32
[9]
Improved accuracy of pedicle screw insertion with computer-assisted surgery - A prospective clinical trial of 30 patients [J].
Laine, T ;
Schlenzka, D ;
Makitalo, K ;
Tallroth, K ;
Nolte, LP ;
Visarius, H .
SPINE, 1997, 22 (11) :1254-1258
[10]
Accuracy of pedicle screw insertion with and without computer assistance: a randomised controlled clinical study in 100 consecutive patients [J].
Laine, T ;
Lund, T ;
Ylikoski, M ;
Lohikoski, J ;
Schlenzka, D .