Advantages using pulse shape discrimination to assign the depth of interaction information (DOI) from a multi layer phoswich detector

被引:41
作者
Schmand, M [1 ]
Eriksson, L
Casey, ME
Wienhard, K
Flügge, G
Nutt, R
机构
[1] CTI PET Syst Inc, Knoxville, TN USA
[2] CTI Inc, Knoxville, TN USA
[3] Max Planck Inst Neurol Res, Cologne, Germany
[4] Rhein Westfal TH Aachen, Inst Phys 3, D-5100 Aachen, Germany
[5] Karolinska Inst, S-10401 Stockholm, Sweden
关键词
D O I
10.1109/23.790808
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Recently new high resolution brain PET and PET/SPECT tomographs have been discussed using phoswich detectors. The detector design is based on two layers of scintillation crystals in the z-axis with different physical properties, such as light decay time or light yield. This makes it feasible to assign the depth of interaction (DOI) information in order to be able to correct for spatial degradation in PET or to operate the detector in PET or SPECT mode. The pixel DOI detectors are cost effectively arranged in a block design [1]. In this work we discuss the possibility of separating events from the two different layers in the time domain using pulse shape discrimination or in the energy domain using pulse height discrimination. Furthermore, we have investigated the feasibility of interdetector scatter suppression using pulse shape discrimination. The measurements have been done using a LSO/LSO high resolution PET detector, a LSO and GSO crystal in coincidence and a NaT(Tl)/LSO combined PET/SPECT detector. The investigations show, that che pulse shape discrimination technique has a significant higher identification probability compared to the pulse height discrimination and does make a interdetector scatter detection feasible.
引用
收藏
页码:985 / 990
页数:6
相关论文
共 7 条
[1]   DESIGN OF A HIGH-RESOLUTION POSITRON EMISSION TOMOGRAPH USING SOLID-STATE SCINTILLATION DETECTORS [J].
CARRIER, C ;
MARTEL, C ;
SCHMITT, D ;
LECOMTE, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) :685-690
[2]   A MULTICRYSTAL 2-DIMENSIONAL BGO DETECTOR SYSTEM FOR POSITRON EMISSION TOMOGRAPHY [J].
CASEY, ME ;
NUTT, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1986, 33 (01) :460-463
[3]   Investigation of LSO crystals for high spatial resolution positron emission tomography [J].
Casey, ME ;
Eriksson, L ;
Schmand, M ;
Andreaco, MS ;
Paulus, M ;
Dahlbom, M ;
Nutt, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :1109-1113
[4]  
MELCHER CL, 1998, IEEE MED IM C REC TO
[5]   A METHOD FOR CORRECTING THE DEPTH-OF-INTERACTION BLURRING IN PET CAMERAS [J].
ROGERS, JG .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1995, 14 (01) :146-150
[6]   Performance results of a new DOI detector block for a High Resolution PET-LSO Research Tomograph HRRT [J].
Schmand, M ;
Eriksson, L ;
Casey, ME ;
Andreaco, MS ;
Melcher, C ;
Wienhard, K ;
Flügge, G ;
Nutt, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (06) :3000-3006
[7]   DEVELOPMENT OF CONTINUOUS DETECTORS FOR A HIGH-RESOLUTION ANIMAL PET SYSTEM [J].
SIEGEL, S ;
CHERRY, SR ;
RICCI, AR ;
SHAO, YP ;
PHELPS, ME .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (04) :1069-1074