DEVELOPMENT OF CONTINUOUS DETECTORS FOR A HIGH-RESOLUTION ANIMAL PET SYSTEM

被引:26
作者
SIEGEL, S
CHERRY, SR
RICCI, AR
SHAO, YP
PHELPS, ME
机构
[1] Crump Institute for Biological Imaging, Dept. of Molecular and Medical Pharmacology UCLA School of Medicine, Los Angeles
关键词
D O I
10.1109/23.467747
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We propose a design for a high resolution, gamma-camera style detector that is suitable for use in a positron emission tomograph dedicated to small animal research. Through Monte Carlo simulation we modeled the performance of a detector composed of one 76.2 x 76.2 x 8 mm thick LSO crystal coupled to a 3 '' square position sensitive photomultiplier tube (PS-PMT). We investigated the effect of optical coupling compounds, surface treatment and depth of interaction on the quantity (efficiency) and distribution (spread) of scintillation photons reaching the photocathode. We also investigated linearization of the position response. We propose a PET system consisting of fourteen of these detectors in 2 rings, yielding a 16 cm diameter by 15 cm long tomograph. It would operate in 3-D mode subtending a 68% solid angle to the center. The expected spatial resolution is less than or equal to 2 mm, with a system efficiency of similar to 10% at the center (200 keV lower threshold) and a singles count rate capability of approximately 10(6) cps per detector.
引用
收藏
页码:1069 / 1074
页数:6
相关论文
共 15 条
[1]   A MULTICRYSTAL 2-DIMENSIONAL BGO DETECTOR SYSTEM FOR POSITRON EMISSION TOMOGRAPHY [J].
CASEY, ME ;
NUTT, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1986, 33 (01) :460-463
[2]  
CHERRY SR, UNPUB IEEE T NUCL SC
[3]  
CUTLER PD, 1992, J NUCL MED, V33, P595
[4]  
HARSHAW QS, SCINTILLATION DETECT
[5]   CAPABILITY EVALUATION OF A SINOGRAM ERROR-DETECTION AND CORRECTION METHOD IN COMPUTED-TOMOGRAPHY [J].
HUANG, SC ;
YU, DC .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1992, 39 (04) :1106-1110
[6]  
KARP JS, 1990, J NUCL MED, V31, P617
[7]  
KEIL G, 1970, NUCL INSTRUM METHODS, V83, P145, DOI [10.1016/0029-554X(70)90550-1, 10.1016/0029-554X(70)90813-X]
[8]   LIGHT COLLECTION IN SCINTILLATION DETECTOR COMPOSITES FOR NEUTRON DETECTION [J].
KNOLL, GF ;
KNOLL, TF ;
HENDERSON, TM .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) :872-875
[9]  
KUME H, 1994, HAMAMATSU PHOTONICS
[10]  
MARRIOTT CJ, 1994, J NUCL MED, V35, P1390