Geometric analysis of light-field position of a multileaf collimator with curved ends

被引:58
作者
Boyer, AL
Li, SD
机构
[1] Department of Radiation Oncology, Stanford University, School of Medicine, Stanford
关键词
multileaf collimator (MLC); field edge; transmission penumbra;
D O I
10.1118/1.597996
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this work was to analyze the difference between the light-field edge and the radiation field edge produced by multileaf collimator (MLC) leaves that are constructed with rounded ends. A formula was derived using three-dimensional geometry predicting the location of the ideal point-source light-field edge relative to the geometric projection of the tip of the rounded leaf end. The theoretical edge of the radiation field for a point source was also calculated along the projection of a chord whose length was one Half-Value Thickness (HVT). Analytic solutions were found for the radiation-field edge and the transmission penumbra. The displacement of the light-field edge relative to the projection of the leaf tip varied from 0 mm for the leaf tip projected onto the central axis to approximately 3.2 mm for the leaf tip projected to 20 cm off the central axis. The light-field edge was always displaced into the unblocked area. The location of the projection of the 1 HVT chord was displaced from the light-field edge by a small and essentially constant distance (approximately 0.3 mm) for all field sizes. The transmission penumbra (80%-20%) was about 1.2 mm for all leaf positions. The calibration of the leaf position must account for the nonlinear relationship between the leaf-tip projection and the light-field (or radiation-field) edge. (C) 1997 American Association of Physicists in Medicine.
引用
收藏
页码:757 / 762
页数:6
相关论文
共 10 条
[1]   CLINICAL DOSIMETRY FOR IMPLEMENTATION OF A MULTILEAF COLLIMATOR [J].
BOYER, AL ;
OCHRAN, TG ;
NYERICK, CE ;
WALDRON, TJ ;
HUNTZINGER, CJ .
MEDICAL PHYSICS, 1992, 19 (05) :1255-1261
[2]  
BOYER AL, 1996, TELETHERAPHY PRESENT
[3]   Dosimetric evaluation of the conformation of the multileaf collimator to irregularly shaped fields [J].
Frazier, A ;
Du, M ;
Wong, J ;
Vicini, F ;
Taylor, R ;
Yu, C ;
Matter, R ;
Martinez, A ;
Yan, D .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 33 (05) :1229-1238
[4]   EVALUATION OF MULTILEAF COLLIMATOR DESIGN FOR A PHOTON-BEAM [J].
GALVIN, JM ;
SMITH, AR ;
MOELLER, RD ;
GOODMAN, RL ;
POWLIS, WD ;
RUBENSTEIN, J ;
SOLIN, LJ ;
MICHAEL, B ;
NEEDHAM, M ;
HUNTZINGER, CJ ;
KLIGERMAN, MM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1992, 23 (04) :789-801
[5]   CHARACTERIZATION OF A MULTILEAF COLLIMATOR SYSTEM [J].
GALVIN, JM ;
SMITH, AR ;
LALLY, B .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 25 (02) :181-192
[6]   DOSIMETRIC CHARACTERISTICS OF A COMMERCIAL MULTILEAF COLLIMATOR [J].
HUQ, MS ;
YU, Y ;
CHEN, ZP ;
SUNTHARALINGAM, N .
MEDICAL PHYSICS, 1995, 22 (02) :241-247
[7]   THE DESIGN AND PERFORMANCE-CHARACTERISTICS OF A MULTILEAF COLLIMATOR [J].
JORDAN, TJ ;
WILLIAMS, PC .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (02) :231-251
[8]   Clinical implementation of a commercial multileaf collimator: Dosimetry, networking, simulation, and quality assurance [J].
Klein, EE ;
Harms, WB ;
Low, DA ;
Willcut, V ;
Purdy, JA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 33 (05) :1195-1208
[9]  
LOSASSO T, 1992, MED PHYS, V19, P816
[10]   DOSE DISTRIBUTIONS OF X-RAY FIELDS AS SHAPED WITH MULTILEAF COLLIMATORS [J].
ZHU, Y ;
BOYER, AL ;
DESOBRY, GE .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (01) :163-173