PDMS-based opto-fluidic micro flow cytometer with two-color, multi-angle fluorescence detection capability using PIN photodiodes

被引:159
作者
Tung, YC
Zhang, M
Lin, CT
Kurabayashi, K
Skerlos, SJ
机构
[1] Univ Michigan, Dept Mech Engn, Ann Arbor, MI 48109 USA
[2] Univ Michigan, Dept Elect Engn & Comp Sci, Ann Arbor, MI 48109 USA
[3] Univ Michigan, Dept Macromol Sci & Engn, Ann Arbor, MI 48109 USA
基金
美国国家科学基金会;
关键词
PDMS observation cell; self-aligned optical fibers; spectral flow cytometry; PIN photodiode; microfluidics; multi-parameter biological detection;
D O I
10.1016/j.snb.2003.10.010
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This research describes the design and performance of a microfluidic observation channel capable of performing flow cytometry measurements using solid-state lasers and PIN-based photodetectors. The observation cell, fabricated using poly(dimethylsiloxane) (PDMS)-based soft lithography, is designed with integrated microgrooves that permit the physical registration of optical fibers that serve as optical waveguides for laser excitation and fluorescence detection. The optical fiber arrangement is designed to permit simultaneous multi-color excitation of sample particles, at a single interrogation point in the observation channel, using multiple angles of excitation and detection. The multiple angles can be used to increase signal-to-noise ratio (SIN) by performing two independent fluorescence measurements simultaneously, and to permit multiple simultaneous cytometric measurements to occur at a single interrogation point. Moreover, the multi-angle micro flow cytometer is proven to provide consistent color differentiation that allows the design to achieve spectrometer capabilities not possible on a conventional flow cytometer. The micro flow cytometer performance was evaluated by considering the detection of fluorescent microspheres (15 mum) and nucleic acid labeled fungus (Saccharomyces cerevisiae). Multiple angles (45degrees, 135degrees, and 180degrees) between fluorescence excitation and detection were investigated with respect to S/N, S/N coefficient of variation, and particle recovery. The experiments demonstrate the feasibility of developing a high S/N single-cell fluorescence detection system based exclusively on silicon-based PIN photodiodes coupled with the lock-in amplification technique. The technique is applicable to the detection of scatter and fluorescent signals emanating from inorganic particles and biological cells, and is highly amenable to economical manufacturing, miniaturization, and low-power operation. Consequently, the design approach and performance results described in this investigation have applicability to the development of practical and low-cost cytometric micro total analysis systems (muTAS). (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:356 / 367
页数:12
相关论文
共 12 条
[1]   An integrated nanoliter DNA analysis device [J].
Burns, MA ;
Johnson, BN ;
Brahmasandra, SN ;
Handique, K ;
Webster, JR ;
Krishnan, M ;
Sammarco, TS ;
Man, PM ;
Jones, D ;
Heldsinger, D ;
Mastrangelo, CH ;
Burke, DT .
SCIENCE, 1998, 282 (5388) :484-487
[2]   Optical particle detection integrated in a dielectrophoretic lab-on-a-chip [J].
Cui, L ;
Zhang, T ;
Morgan, H .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2002, 12 (01) :7-12
[3]   Methodology for detection and typing of foodborne microorganisms [J].
de Boer, E ;
Beumer, RR .
INTERNATIONAL JOURNAL OF FOOD MICROBIOLOGY, 1999, 50 (1-2) :119-130
[4]  
Givan, 2001, FLOW CYTOMETRY 1 PRI, DOI 10.1002/0471223948
[5]  
HUN D, 2002, BIOMED MICRODEVICES, V4, P141
[6]   The mechanical properties of the rubber elastic polymer polydimethylsiloxane for sensor applications [J].
Lotters, JC ;
Olthuis, W ;
Veltink, PH ;
Bergveld, P .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 1997, 7 (03) :145-147
[7]   Polydimethylsiloxane as an elastic material applied in a capacitive accelerometer [J].
Lotters, JC ;
Olthuis, W ;
Veltink, PH ;
Bergveld, P .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 1996, 6 (01) :52-54
[8]  
NAMASIVAYAM V, 2001, P 5 INT S MICR TOT A, P355
[9]  
RO KW, 2001, P 5 INT S MICR TOT A
[10]   Integrated diode detector and optical fibres for in situ detection within micromachined polymerase chain reaction chips [J].
Schabmueller, CGJ ;
Pollard, JR ;
Evans, AGR ;
Wilkinson, JS ;
Ensell, G ;
Brunnschweiler, A .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2001, 11 (04) :329-333