Nine commercially available silica based, wide-pore columns for protein separation were tested and compared using some low molecular weight aromatic test compounds recommended in the literature. From the retention values obtained under identical mobile phase conditions the strength of stationary phases could have been well characterised by spectral mapping analysis (SMA). However, inconsistency with experience was observed with respect to the polar-neutral and polar-ionic nature of packings. After equalising the mobile phase conditions, the confounding effect of the different column strength could have been eliminated and the selectivity differences amongst the stationary phases could have been revealed.
引用
收藏
页码:1667 / 1687
页数:21
相关论文
共 18 条
[1]   Characterization of wide-pore reversed-phase columns for biopolymer separations .1. Single-parametric evaluation [J].
Bede, A ;
Rippel, G ;
Szepesy, L ;
Claessens, HA .
JOURNAL OF CHROMATOGRAPHY A, 1996, 728 (1-2) :179-188
[2]  
CLAESSENS HA, 1993, LC GC EUR, V6, P692
[3]   CHEMOMETRIC CLASSIFICATION OF REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY COLUMNS [J].
DELANEY, MF ;
PAPAS, AN ;
WALTERS, MJ .
JOURNAL OF CHROMATOGRAPHY, 1987, 410 (01) :31-41
[4]   CHROMATOGRAPHIC CHARACTERIZATION OF SILICA-BASED REVERSED PHASES [J].
ENGELHARDT, H ;
LOW, H ;
GOTZINGER, W .
JOURNAL OF CHROMATOGRAPHY, 1991, 544 (1-2) :371-379
[5]   COMPARISON AND CHARACTERIZATION OF REVERSED PHASES [J].
ENGELHARDT, H ;
JUNGHEIM, M .
CHROMATOGRAPHIA, 1990, 29 (1-2) :59-68
[6]   COMPARISON OF COLUMNS FOR REVERSED-PHASE LIQUID-CHROMATOGRAPHY [J].
GOLDBERG, AP .
ANALYTICAL CHEMISTRY, 1982, 54 (02) :342-345
[7]   SPECTRAL MAPPING ANALYSIS - A METHOD FOR THE CHARACTERIZATION OF STATIONARY PHASES [J].
HAMOIR, T ;
SANCHEZ, FC ;
BOURGUIGNON, B ;
MASSART, DL .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1994, 32 (11) :488-498
[8]   CHEMOMETRIC CATEGORIZATION OF OCTADECYLSILYL BONDED-PHASE SILICA COLUMNS USING TEST MIXTURES AND CONFIRMATION OF RESULTS WITH PHARMACEUTICAL COMPOUND SEPARATIONS [J].
OLSEN, BA ;
SULLIVAN, GR .
JOURNAL OF CHROMATOGRAPHY A, 1995, 692 (1-2) :147-159
[9]   FACTOR-ANALYSIS AND EXPERIMENT DESIGN IN HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY .10. CHEMOMETRIC CHARACTERIZATION OF PACKINGS, SOLVENTS AND SOLUTES WITH HIERARCHICAL ASCENDING CLASSIFICATION AND CORRESPONDENCE FACTOR-ANALYSIS [J].
RIGHEZZA, M ;
CHRETIEN, JR .
JOURNAL OF CHROMATOGRAPHY, 1991, 544 (1-2) :393-411
[10]   HYDROPHOBIC INTERACTION CHROMATOGRAPHY OF PROTEINS ON AN ALKYL-SUPEROSE COLUMN [J].
RIPPEL, G ;
SZEPESY, L .