Dynamic torque and apical forces of ProFile .04 rotary instruments during preparation of curved canals

被引:154
作者
Peters, OA [1 ]
Barbakow, F [1 ]
机构
[1] Univ Zurich, Dept Prevent Dent Cariol & Periodontol, Div Endodontol, CH-8028 Zurich, Switzerland
关键词
curved canals; cyclic fatigue; dynamic torque; force; Ni-Ti;
D O I
10.1046/j.0143-2885.2001.00494.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Aim To analyse torque and force generated whilst shaping curved canals using rotary instruments. Methodology A specially designed computer-controlled testing platform was used to record events during the shaping of straight and curved canals in plastic blocks and in extracted human teeth using ProFile .04 instruments. Size 40 apical stops were prepared using crown-down, apical preparation and step-back procedures. Maximum torque, apically directed force and the numbers of revolutions were recorded at a resolution of 100 samples s(-1) . Load causing separation as required by the ISO 3630-1 test and cyclic fatigue was also recorded. Mean maximum scores were calculated and statistically tested using one- and two-way analyses of variance. Results Highest and lowest torque scores were recorded, respectively, in straight canals in plastic blocks at 25 Nmm and in natural canals at 14 Nmm. Significant differences were recorded for canal type and preparation phase (P < 0.0001). Loads causing separation varied from 3.7 to 32.3 Nmm. Apically directed forces ranged from 1 to 7.5 N. Again, there were significant differences depending on canal type and preparation phase (P < 0.0001). The number of revolutions during preparation ranged from 18 to 41. Size 15, 30 and 45 ProFile .04 instruments separated after 581, 430 and 402 revolutions, respectively, in a standard cyclic fatigue test. Conclusions The new torque-testing platform details physical parameters during preparation of curved canals. To improve predictability, instrumentation sequences must be tested for excessively high torsional moments or forces. This study indicated that up to 10 curved canals could be safely prepared with a sequence of ProFile. 04 rotary instruments without separation due to cyclic fatigue. Efforts should continue to correlate root canal anatomy with torque and force generated during rotary root canal preparation.
引用
收藏
页码:379 / 389
页数:11
相关论文
共 38 条
[1]   The 'Lightspeed' preparation technique evaluated by Swiss clinicians after attending continuing education courses [J].
Barbakow, F ;
Lutz, F .
INTERNATIONAL ENDODONTIC JOURNAL, 1997, 30 (01) :46-50
[2]   Analysis of forces developed during root canal preparation with the balanced force technique [J].
Blum, JY ;
Machtou, P ;
Esber, S ;
Micallef, JP .
INTERNATIONAL ENDODONTIC JOURNAL, 1997, 30 (06) :386-396
[3]   Location of contact areas on rotary Profile® instruments in relationship to the forces developed during mechanical preparation on extracted teeth [J].
Blum, JY ;
Machtou, P ;
Micallef, JP .
INTERNATIONAL ENDODONTIC JOURNAL, 1999, 32 (02) :108-114
[4]   Analysis of forces developed during mechanical preparation of extracted teeth using Profile NiTi rotary instruments [J].
Blum, JY ;
Cohen, A ;
Machtou, P ;
Micallef, JP .
INTERNATIONAL ENDODONTIC JOURNAL, 1999, 32 (01) :24-31
[5]  
Bryant ST, 1998, INT ENDOD J, V31, P275
[6]  
Bryant ST, 1998, INT ENDOD J, V31, P282
[7]   TORSIONAL AND STIFFNESS PROPERTIES OF CANAL MASTER-U STAINLESS-STEEL AND NITINOL INSTRUMENTS [J].
CAMPS, J ;
PERTOT, WJ .
JOURNAL OF ENDODONTICS, 1994, 20 (08) :395-398
[8]   TORSIONAL PROPERTIES OF STAINLESS-STEEL CANAL MASTER-U AND FLEXOGATES [J].
CAMPS, JJ ;
PERTOT, WJ .
INTERNATIONAL ENDODONTIC JOURNAL, 1994, 27 (06) :334-338
[9]  
CAMPS JJ, 1994, ENDOD DENT TRAUMATOL, V10, P260
[10]   Wear of nickel-titanium lightspeed instruments evaluated by scanning electron microscopy [J].
Eggert, C ;
Peters, O ;
Barbakow, F .
JOURNAL OF ENDODONTICS, 1999, 25 (07) :494-497