Rise in chronic defibrillation energy requirements necessitating implantable defibrillator lead system revision

被引:7
作者
Daoud, EG [1 ]
Man, KC [1 ]
Morady, F [1 ]
Strickberger, SA [1 ]
机构
[1] UNIV MICHIGAN HOSP,DEPT INTERNAL MED,DIV CARDIOL,ANN ARBOR,MI 48109
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1997年 / 20卷 / 03期
关键词
defibrillation energy; implantable defibrillator;
D O I
10.1111/j.1540-8159.1997.tb03890.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The chronic defibrillation energy requirement (DER) is believed to remain clinically stable in patients with defibrillators. Six patients (two with an epicardial and four with a nonthoracotomy system) were identified with a rise in their chronic DER, which eliminated a 10-J safety margin, thus necessitating a defibrillator lead system revision. The mean increase in DER was 14.7 +/- 4J and was discovered at a mean of 16.0 +/- 18 months (range 2-41) following implantation. Management included placement of a defibrillator with a biphasic waveform, placement of an additional defibrillation electrode, or both. At 2 months following revision of the defibrillation system, a 10-J DER safety margin was present in each patient. In some patients, there is a progressive increase in the chronic DER with elimination of a 10-J safety margin necessitating revision of the defibrillation system. Routine reevaluation of the chronic DER, therefore, is necessary to identify these patients.
引用
收藏
页码:714 / 719
页数:6
相关论文
共 16 条
[1]  
BARDY GH, 1989, AM J CARDIOL, V60, P618
[2]   EFFECTS OF PROCAINAMIDE AND LIDOCAINE ON DEFIBRILLATION ENERGY-REQUIREMENTS IN PATIENTS RECEIVING IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR DEVICES [J].
ECHT, DS ;
GREMILLION, ST ;
LEE, JT ;
RODEN, DM ;
MURRAY, KT ;
BORGANELLI, M ;
CRAWFORD, DM ;
STEWART, JR ;
HAMMON, JW .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1994, 5 (09) :752-760
[3]   COMPLICATIONS OF AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATORS - RADIOGRAPHIC, CT, AND ECHOCARDIOGRAPHIC EVALUATION [J].
GOODMAN, LR ;
ALMASSI, GH ;
TROUP, PJ ;
GURNEY, JW ;
VESETHROGERS, J ;
CHAPMAN, PD ;
WETHERBEE, JN .
RADIOLOGY, 1989, 170 (02) :447-452
[4]  
LANG DJ, 1994, IMPLANTABLE CARDIOVE, P223
[5]  
LEMMER JH, 1993, CIRCULATION, V88, P113
[6]   PRELIMINARY EXPERIENCE WITH A HYBRID NONTHORACOTOMY DEFIBRILLATING SYSTEM THAT INCLUDES A BIPHASIC DEVICE - COMPARISON WITH A STANDARD MONOPHASIC DEVICE USING THE SAME LEAD SYSTEM [J].
NATALE, A ;
SRA, J ;
AXTELL, K ;
MAGLIO, C ;
DHALA, A ;
DESHPANDE, S ;
JAZAYERI, M ;
WASE, A ;
AKHTAR, M .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 24 (02) :406-412
[7]   EFFECT OF BIPHASIC WAVE-FORM PULSE ON ENDOCARDIAL DEFIBRILLATION EFFICACY IN HUMANS [J].
NEUZNER, J ;
PITSCHNER, HF ;
HUTH, C ;
SCHLEPPER, M .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (02) :207-212
[8]   LIMITATIONS AND LATE COMPLICATIONS OF 3RD-GENERATION AUTOMATIC CARDIOVERTER-DEFIBRILLATORS [J].
NUNAIN, SO ;
ROELKE, M ;
TROUTON, T ;
OSSWALD, S ;
KIM, YH ;
SOSASUAREZ, G ;
BROOKS, DR ;
MCGOVERN, B ;
GUY, M ;
TORCHIANA, DF ;
VLAHAKES, GJ ;
GARAN, H ;
RUSKIN, JN .
CIRCULATION, 1995, 91 (08) :2204-2213
[9]  
RIGDEN LB, 1995, J AM COLL CARDIOL, V25, pA85
[10]   POSTOPERATIVE LEAD-RELATED COMPLICATIONS IN PATIENTS WITH NONTHORACOTOMY DEFIBRILLATION LEAD SYSTEMS [J].
SCHWARTZMAN, D ;
NALLAMOTHU, N ;
CALLANS, DJ ;
PREMINGER, MW ;
GOTTLIEB, CD ;
MARCHLINSKI, FE .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1995, 26 (03) :776-786