Practical approach to electron beam dosimetry at extended SSD

被引:23
作者
Cygler, J
Li, XA
Ding, GX
Lawrence, E
机构
[1] Department of Medical Physics, Ottawa Regional Cancer Centre, Civic Division, Ottawa, Ont. K1Y 4K7
关键词
D O I
10.1088/0031-9155/42/8/003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper presents a practical approach to clinical dosimetry for electron beams at extended source-to-surface distance (SSD). Characteristics of electron beams from a Siemens MD-2 accelerator are presented at a nominal SSD of 100 cm and an extended SSD of 115 cm. Relative output factors are measured at both 100 and 115 cm SSDs for a range of square field sizes for each beam energy. The change of output with SSD does not follow the inverse square law, if the nominal SSD is used. This deviation is larger for lower beam energy and smaller field sizes. An effective SSD, SSDeff, has been determined at d(max) for each beam energy as a function of field size. A comprehensive approach to treatment time calculations is proposed.
引用
收藏
页码:1505 / 1514
页数:10
相关论文
共 11 条
[1]  
[Anonymous], 1984, 35 ICRU
[2]   ELECTRON-BEAM CHARACTERISTICS AT EXTENDED TREATMENT DISTANCES [J].
DAS, IJ ;
MCGEE, KP ;
CHENG, CW .
MEDICAL PHYSICS, 1995, 22 (10) :1667-1674
[3]   A MODEL FOR ELECTRON-BEAM APPLICATOR SCATTER [J].
EBERT, MA ;
HOBAN, PW .
MEDICAL PHYSICS, 1995, 22 (09) :1419-1429
[4]   DETERMINATION OF THE SOURCE POSITION FOR THE ELECTRON-BEAMS FROM A HIGH-ENERGY LINEAR-ACCELERATOR [J].
JAMSHIDI, A ;
KUCHNIR, FT ;
REFT, CS .
MEDICAL PHYSICS, 1986, 13 (06) :942-948
[5]   EFFECT OF AIR SPACE ON DEPTH DOSE IN ELECTRON-BEAM THERAPY [J].
KHAN, FM ;
SEWCHAND, W ;
LEVITT, SH .
RADIOLOGY, 1978, 126 (01) :249-251
[6]   CLINICAL ELECTRON-BEAM DOSIMETRY - REPORT OF AAPM RADIATION-THERAPY COMMITTEE TASK GROUP NO-25 [J].
KHAN, FM ;
DOPPKE, KP ;
HOGSTROM, KR ;
KUTCHER, GJ ;
NATH, R ;
PRASAD, SC ;
PURDY, JA ;
ROZENFELD, M ;
WERNER, BL .
MEDICAL PHYSICS, 1991, 18 (01) :73-109
[7]   ELECTRON MASS SCATTERING POWERS - MONTE-CARLO AND ANALYTICAL CALCULATIONS [J].
LI, XA ;
ROGERS, DWO .
MEDICAL PHYSICS, 1995, 22 (05) :531-541
[8]   Effective SSD for electron beams as a function of energy and beam collimation [J].
Roback, DM ;
Khan, FM ;
Gibbons, JP ;
Sethi, A .
MEDICAL PHYSICS, 1995, 22 (12) :2093-2095
[9]  
Saw C B, 1994, Med Dosim, V19, P77
[10]   DOSE DISTRIBUTION CONSIDERATIONS OF MEDIUM-ENERGY ELECTRON-BEAMS AT EXTENDED SOURCE-TO-SURFACE DISTANCE [J].
SHAW, CB ;
AYYANGAR, KM ;
PAWLICKI, T ;
KORB, LJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 32 (01) :159-164