Instrumentation of the ESRF medical imaging facility

被引:97
作者
Elleaume, H
Charvet, AM
Berkvens, P
Berruyer, G
Brochard, T
Dabin, Y
Dominguez, MC
Draperi, A
Fiedler, S
Goujon, G
Le Duc, G
Mattenet, M
Nemoz, C
Perez, M
Renier, M
Schulze, C
Spanne, P
Suortti, P
Thomlinson, W
Esteve, F
Bertrand, B
Le Bas, JF
机构
[1] Univ J Fourier, Unite IRM, CHU, F-38043 Grenoble, France
[2] European Synchrotron Radiat Facil, F-38043 Grenoble, France
[3] Paul Scherrer Inst, Swiss Light Source, CH-5232 Villigen, Switzerland
[4] Brookhaven Natl Lab, NSLS, Upton, NY 11973 USA
关键词
synchrotron; medical imaging; monochromatic beams; angiography; computed tomography; germanium detector;
D O I
10.1016/S0168-9002(99)00167-9
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
At the European Synchrotron Radiation Facility (ESRF) a beamport has been instrumented for medical research programs. Two facilities have been constructed for alternative operation. The first one is devoted to medical imaging and is focused on intravenous coronary angiography and computed tomography (CT). The second facility is dedicated to pre-clinical microbeam radiotherapy (MRT). This paper describes the instrumentation for the imaging facility. Two monochromators have been designed, both are based on bent silicon crystals in the Laue geometry. A versatile scanning device has been built for pre-alignment and scanning of the patient through the X-ray beam in radiography or CT modes. An intrinsic germanium detector is used together with large dynamic range electronics (16 bits) to acquire the data. The beamline is now at the end of its commissioning phase; intravenous coronary angiography is intended to start in 1999 with patients and the CT pre-clinical program is underway on small animals. The first in vivo images obtained on animals in angiography and CT modes are presented to illustrate the performances of these devices. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:513 / 527
页数:15
相关论文
共 20 条
  • [1] Diffraction enhanced x-ray imaging
    Chapman, D
    Thomlinson, W
    Johnston, RE
    Washburn, D
    Pisano, E
    Gmur, N
    Zhong, Z
    Menk, R
    Arfelli, F
    Sayers, D
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (11) : 2015 - 2025
  • [2] CHARVET AM, 1997, MED APPL SYNCHROTRON
  • [3] INTRAVENOUS CORONARY ANGIOGRAPHY WITH SYNCHROTRON-RADIATION
    DIX, WR
    [J]. PROGRESS IN BIOPHYSICS & MOLECULAR BIOLOGY, 1995, 63 (02) : 159 - 191
  • [4] DOLBNYA IP, 1990, PHYS MEDICA, V6, P313
  • [5] Feldmann HJ, 1998, MED RAD DIA IMG, P47
  • [6] COMPLICATIONS ASSOCIATED WITH CARDIAC-CATHETERIZATION AND ANGIOGRAPHY
    KENNEDY, JW
    [J]. CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1982, 8 (01): : 5 - 11
  • [7] LABICHE JC, 1994, Patent No. 9413614
  • [8] MONOCHROMATIC COMPUTED-TOMOGRAPHY OF THE HUMAN BRAIN USING SYNCHROTRON X-RAYS - TECHNICAL FEASIBILITY
    NACHALIEL, E
    DILMANIAN, FA
    GARRETT, RF
    THOMLINSON, WC
    CHAPMAN, LD
    GMUR, NF
    LAZARZ, NM
    MOULIN, HR
    RIVERS, ML
    RARBACK, H
    STEFAN, PM
    SPANNE, P
    LUKE, PN
    PEHL, R
    THOMPSON, AC
    MILLER, M
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1992, 319 (1-3) : 305 - 310
  • [9] CORONARY-ARTERIES - NEW DESIGN FOR 3-DIMENSIONAL ARTERIAL PHANTOMS
    RENAUDIN, CP
    BARBIER, B
    RORIZ, R
    REVEL, D
    AMIEL, M
    [J]. RADIOLOGY, 1994, 190 (02) : 579 - 582
  • [10] The heart of the matter: Noninvasive coronary artery imaging
    Rogers, LF
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1998, 170 (04) : 841 - 841