Novel high-resolution temperature probe for radiofrequency dosimetry

被引:9
作者
Schuderer, J [1 ]
Schmid, T
Urban, G
Samaras, T
Kuster, N
机构
[1] Swiss Fed Inst Technol, Fdn Res Informat Technol Soc ITIS, Integrated Syst Lab, CH-8092 Zurich, Switzerland
[2] Schmid & Partner Engn AG, CH-8004 Zurich, Switzerland
[3] Univ Freiburg, IMTEK, D-79110 Freiburg, Germany
[4] Aristotle Univ Thessaloniki, Dept Phys, GR-54124 Thessaloniki, Greece
关键词
D O I
10.1088/0031-9155/49/6/N04
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A novel integrated thermistor probe for temperature evaluations in radiofrequency-heated environments was realized. The probe's sensitive area is based on a highly resistive 50 gm x 100 mum layer of amorphous germanium processed on a glass tip. The small dimensions allow measurements with a distance as close as 150 mum from solid boundaries. Due to its high temperature resolution of 4 mK and its short response time of the order of 10 ms, the sensor is very well suited for dosimetric measurements in strong absorption gradients. The influence of radiofrequency (RF) electric fields on the signal is minimized due to the high resistance of the sensor and the leads. The probe was successfully used to determine the highly nonuniform absorption distribution resulting from the RF exposure of cell cultures placed in Petri dishes.
引用
收藏
页码:N83 / N92
页数:10
相关论文
共 12 条
[1]   PROBE FOR MEASURING TEMPERATURE IN RADIO-FREQUENCY-HEATED MATERIAL [J].
BOWMAN, RR .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1976, 24 (01) :43-45
[2]  
Burkhardt M, 1996, BIOELECTROMAGNETICS, V17, P483, DOI 10.1002/(SICI)1521-186X(1996)17:6<483::AID-BEM8>3.0.CO
[3]  
2-#
[4]  
KUTTNER H, 1991, SENSOR ACTUAT A-PHYS, V25, P641
[5]  
MICHALSKI L, 2001, TEMPERATURE MEASUREM, P284
[6]   On the assumption of negligible heat diffusion during the thermal measurement of a nonuniform specific absorption rate [J].
Moros, EG ;
Pickard, WF .
RADIATION RESEARCH, 1999, 152 (03) :312-320
[7]   Millimeter-resolution E-field probe for isotropic measurement in lossy media between 100 MHz and 20 GHz [J].
Pokovic, K ;
Schmid, T ;
Kuster, N .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2000, 49 (04) :873-878
[8]   Electromagnetic and heat transfer computations for non-ionizing radiation dosimetry [J].
Samaras, T ;
Regli, P ;
Kuster, N .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (08) :2233-2246
[9]   Automated E-field scanning system for dosimetric assessments [J].
Schmid, T ;
Egger, O ;
Kuster, N .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1996, 44 (01) :105-113
[10]  
SCHUDERER J, 2004, UNPUB IEEE T MICROW