Ultra-low-volume, real-time measurements of lactate from the single heart cell using microsystems technology

被引:80
作者
Cai, XX
Klauke, N
Glidle, A
Cobbold, P
Smith, GL
Cooper, JM [1 ]
机构
[1] Univ Glasgow, Dept Elect & Elect Engn, Bioelect Res Ctr, Glasgow G12 8LT, Lanark, Scotland
[2] Univ Liverpool, Dept Human Anat & Cell Biol, Liverpool L69 3BX, Merseyside, England
[3] Univ Glasgow, Inst Biomed & Life Sci, Glasgow G12 8QQ, Lanark, Scotland
关键词
D O I
10.1021/ac010941+
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The fabrication of microelectrodes integrated within ultralow-volume microtiter chambers for the amperometric determination of metabolites continues to be of interest in the subject of single-cell and high-throughput screening. The microsystem described in this paper consists of a two-microelectrode sensor with a microfluidic dispensation technology, which is able to deliver both very low titers (6.5 pL) and single heart cells into a low-volume micro-photoelectrochemical cell. Devices were fabricated using photolithography and liftoff giving reproducible sensors integrated within high aspect ratio titer chambers (with a volume of 360 pL, made of the photoepoxy SU8. In this paper, the determination of lactate was optimized using an enzyme-linked assay based upon lactate oxidase, involving the amperometric determination of hydrogen peroxide at +640 mV versus an internal Ag/AgCl pseudoreference. The microsystem (including the microfluidic dispensers and structures as well as the microsensor) was subsequently used to measure the lactate content of single heart cells. Dynamic electrochemical measurements of lactate during cell permeabilization are presented. We also show the use of respiratory uncouplers to simulate ischemia in the single myocyte and show that, as expected, the rate of lactate production from the hypoxic heart cell is greater than that within the normoxic healthy myocyte.
引用
收藏
页码:908 / 914
页数:7
相关论文
共 25 条
[21]   An integrated three-electrode system with a micromachined liquid-junction Ag/AgCl reference electrode [J].
Suzuki, H ;
Hirakawa, T ;
Sasaki, S ;
Karube, I .
ANALYTICA CHIMICA ACTA, 1999, 387 (01) :103-112
[22]   Spatio-temporal resolution of exocytosis from individual cells [J].
Travis, ER ;
Wightman, RM .
ANNUAL REVIEW OF BIOPHYSICS AND BIOMOLECULAR STRUCTURE, 1998, 27 :77-103
[23]   Amperometric biosensors for clinical and therapeutic drug monitoring: a review [J].
Wang, J .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1999, 19 (1-2) :47-53
[24]   KINETICS OF THE SARCOLEMMAL LACTATE CARRIER IN SINGLE HEART-CELLS USING BCECF TO MEASURE PH(I) [J].
WANG, XM ;
LEVI, AJ ;
HALESTRAP, AP .
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 1994, 267 (05) :H1759-H1769
[25]   Simultaneous detection of catecholamine exocytosis and Ca2+ release from single bovine chromaffin cells using a dual microsensor [J].
Xin, Q ;
Wightman, RM .
ANALYTICAL CHEMISTRY, 1998, 70 (09) :1677-1681