Factorial optimization of data acquisition factors for lead isotope ratio determination by inductively coupled plasma mass spectrometry

被引:50
作者
Quetel, CR
Thomas, B
Donard, OFX
Grousset, FE
机构
[1] UNIV BORDEAUX 1,PHOTOPHYS & PHOTOCHIM MOL LAB,URA CNRS 348,F-33405 TALENCE,FRANCE
[2] UNIV BORDEAUX 1,DEPT GEOL & OCEANOG,URA CNRS 197,F-33405 TALENCE,FRANCE
关键词
inductively coupled plasma mass spectrometry; lead isotope ratio; instrumental noise; optimization;
D O I
10.1016/S0584-8547(96)01587-X
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
A factorial optimization of data acquisition factors was performed to minimize non-random instrumental noise on an inductively coupled plasma mass spectrometer. Results have identified instrumental and analytical operating conditions under which lead isotope ratios can be measured precisely and accurately in natural samples. Precision was checked periodically over weeks and ranged 0.10-0.30, 0.15-0.35, 0.15-0.35, 0.10-0.30 and 0.10-0.20 (%) for 206/204, 207/204, 208/204, 206/207 and 208/206 ratios respectively. Mass bias was minimized and the correction factors were typically of 0.6% for 206/204, 1.5% for 207/204, 1.8% for 208/204, 0.8% for 206/207 and 1.2% for 208/206. These results were performed with a dwell time of 20 ms, 40 sweeps per replicate, 8 replicates by analysis, a time factor of 15 applied on mass 204 and three points per peak in the low resolution mode. The total duration time of an analysis was 10:50 min and it was found that optimum concentration of lead in solution was 50 ng ml(-1) on our instrument. (C) 1997 Elsevier Science B.V.
引用
收藏
页码:177 / 187
页数:11
相关论文
共 29 条
[1]   DETERMINATION OF TRACE-METALS IN AN OPEN OCEAN WATER REFERENCE MATERIAL BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
BEAUCHEMIN, D ;
MCLAREN, JW ;
MYKYTIUK, AP ;
BERMAN, SS .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1988, 3 (02) :305-308
[2]   OCCURRENCE AND REDUCTION OF NOISE IN INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY FOR ENHANCED PRECISION IN ISOTOPE RATIO MEASUREMENT [J].
BEGLEY, IS ;
SHARP, BL .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (03) :171-176
[3]   ACCURATE AND PRECISE DETERMINATION OF LEAD ISOTOPE RATIOS IN CLINICAL AND ENVIRONMENTAL-SAMPLES USING INDUCTIVELY COUPLED PLASMA SOURCE-MASS SPECTROMETRY [J].
CAMPBELL, MJ ;
DELVES, HT .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1989, 4 (02) :235-236
[4]  
Cochran W.G. G.M. Cox., 1957, Experimental Design
[5]   NOISE POWER SPECTRAL CHARACTERISTICS OF AN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETER [J].
CRAIN, JS ;
HOUK, RS ;
ECKELS, DE .
ANALYTICAL CHEMISTRY, 1989, 61 (06) :606-612
[6]   STUDIES IN THE DETERMINATION OF LEAD ISOTOPE RATIOS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
DATE, AR ;
CHEUNG, YY .
ANALYST, 1987, 112 (11) :1531-1540
[7]   SELECTION OF MODE FOR THE MEASUREMENT OF LEAD ISOTOPE RATIOS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY AND ITS APPLICATION TO MILK POWDER ANALYSIS [J].
DEAN, JR ;
EBDON, L ;
MASSEY, R .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1987, 2 (04) :369-374
[8]   NOISE CHARACTERISTICS OF AN INDUCTIVELY COUPLED PLASMA-MASS SPECTROMETER [J].
FURUTA, N ;
MONNIG, CA ;
YANG, PY ;
HIEFTJE, GM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1989, 44 (07) :649-656
[9]   OPTIMIZATION OF THE MASS SCANNING RATE FOR THE DETERMINATION OF LEAD ISOTOPE RATIOS USING AN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETER [J].
FURUTA, N .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1991, 6 (03) :199-203
[10]   INTERLABORATORY COMPARISON STUDY ON LEAD ISOTOPE RATIOS DETERMINED BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
FURUTA, N .
ANALYTICAL SCIENCES, 1991, 7 (05) :823-826