The internal consistency of the 'analyst' diffusive sampler - A long-term field test

被引:23
作者
Bertoni, G [1 ]
Tappa, R [1 ]
Allegrini, I [1 ]
机构
[1] Ist Inquinamento Atmosfer, CNR AdR Roma, I-00016 Monterotondo, RM, Italy
关键词
gas chromatography; air monitoring; diffusive sampling; benzene; internal consistency; volatile aromatic compounds;
D O I
10.1007/BF02492194
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A recently developed diffusive sampler, intended for use in long-term ambient measurements of volatile aromatic hydrocarbons, has been tested to determine the upper limits of its Sampling capacity A test has been performed over 12-month period to evaluate these limits in terms of its 'internal consistency (i.e. checking if the sum of the amounts of a given pollutant, determined over shorter consecutive periods, matches the amount determined from a single long-term exposure). Results demonstrate that the uptake rate of the device, in a real (indoor) atmosphere, is constant for approximately 8 months for benzene and for approximately 10 months for toluene. The upper limit for higher homologous species was not reached after exposure for a year to typical ambient concentrations. This means the device can be successfully used for several months, thereby enabling low-cost determinations of ambient BTX concentrations.
引用
收藏
页码:653 / 657
页数:5
相关论文
共 11 条
[1]  
[Anonymous], 1999, Official Journal of the European Communities
[2]   Ambient volatile organic compound monitoring by diffusive sampling. Compatibility of high uptake rate samplers with thermal desorption [J].
Bates, M ;
Gonzalez-Flesca, N ;
Cocheo, V ;
Sokhi, R .
ANALYST, 1997, 122 (12) :1481-1484
[3]   CRITICAL PARAMETERS FOR THE ADSORPTION OF GASEOUS-POLLUTANTS ON PASSIVE SAMPLERS MADE OF LOW SPECIFIC AREA ADSORBENTS [J].
BERTONI, G ;
PERRINO, C ;
FRATARCANGELI, R ;
LIBERTI, A .
ANALYTICAL LETTERS PART A-CHEMICAL ANALYSIS, 1985, 18 (04) :429-438
[4]  
Bertoni G, 2000, ANN CHIM-ROME, V90, P249
[5]  
HARPER M, 1987, AM IND HYG ASSOC J, V48, P214, DOI 10.1202/0002-8894(1987)048<0214:DSAR>2.0.CO
[6]  
2
[7]  
MATTEWS TW, 1987, INDOOR AIR 87, V1, P174
[8]  
POSNER JC, 1985, AM IND HYG ASSOC J, V46, P277, DOI 10.1202/0002-8894(1985)046<0277:ATTOPM>2.3.CO
[9]  
2
[10]   CONVECTIVE-TRANSPORT IN DIFFUSIVE SAMPLERS [J].
UNDERHILL, DW .
TALANTA, 1993, 40 (06) :785-790