Comparison of fiber optics and standard nickel catalytic probes for determination of neutral oxygen atoms concentration

被引:60
作者
Poberaj, I
Mozetic, M
Babic, D
机构
[1] Univ Ljubljana, Fac Math & Phys, Ljubljana 1000, Slovenia
[2] Inst Surface Engn & Optoelect, Ljubljana 1000, Slovenia
来源
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS | 2002年 / 20卷 / 01期
关键词
Neutral oxygen atoms - Nickel catalytic probes - Plasma postglow chamber;
D O I
10.1116/1.1427893
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
The density of neutral oxygen atoms in a plasma postglow chamber was measured with a standard Ni catalytic probe and a fiber optical catalytic probe. A highly dissociated oxygen plasma was created in the discharge chamber with an inductively coupled rf generator with a frequency of 27.12 MHz and output power of 200 W. The measurements were performed in the postglow chamber where a movable recombinator for oxygen atoms was placed. The recombinator enabled the adjustment of O density independently of discharge parameters. The density of neutral oxygen atoms was determined at different pressures and different positions of the recombinator. The O density was between 5 x 10(20) and 2 x 10(21) m(-3). The fiber probe expressed excellent repeatability as compared to the standard probe. As long as the temperature of the catalyst was above 100 degreesC both probes gave similar results. At a lower temperature of the standard probe, however, the optical fiber probe gave more reasonable results. The discrepancy was attributed to the decrease of the catalytic activity of nickel at lower temperatures. (C) 2002 American Vacuum Society.
引用
收藏
页码:189 / 193
页数:5
相关论文
共 21 条
[11]  
Mozetic M, 1998, INFORM MIDEM, V28, P175
[12]   Excited species in H2, N2, O2 microwave flowing discharges and post-discharges [J].
Ricard, A ;
Gaillard, M ;
Monna, V ;
Vesel, A ;
Mozetic, M .
SURFACE & COATINGS TECHNOLOGY, 2001, 142 :333-336
[13]  
Ricard A., 1996, REACTIVE PLASMAS
[14]   Sample collection and preparation for analysis of airborne particulate matter by total reflection X-ray fluorescence spectrometry [J].
Schmeling, M ;
Klockow, D .
ANALYTICA CHIMICA ACTA, 1997, 346 (01) :121-126
[15]  
SIMPSON LA, 1979, AUST OCCA P NEWS, V16, P1
[16]   ATOMIC OXYGEN-METAL SURFACE STUDIES AS APPLIED TO MASS SPECTROMETER MEASUREMENTS OF UPPER PLANETARY ATMOSPHERES. [J].
Sjolander, Gary W. .
1600, (81)
[17]  
Sorli I, 1996, INFORM MIDEM, V26, P35
[18]   Determination of atomic oxygen density with a nickel catalytic probe [J].
Sorli, I ;
Rocak, R .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS, 2000, 18 (02) :338-342
[19]   Behaviour of catalytic probe during surface activation of polyether sulphone [J].
Vesel, A ;
Mozetic, M .
VACUUM, 2001, 61 (2-4) :373-377
[20]  
WERTHEIMER MR, 1998, PLASMA SOURCES POLYM