Optimal phosphor thickness for portal imaging

被引:40
作者
Bissonnette, JP
Cunningham, IA
Munro, P
机构
[1] LONDON REG CANC CTR,LONDON,ON N6A 4L6,CANADA
[2] UNIV WESTERN ONTARIO,DEPT MED BIOPHYS,LONDON,ON N6A 5C1,CANADA
[3] JOHN P ROBARTS RES INST,IMAGING RES LABS,LONDON,ON N6A 5K8,CANADA
[4] VICTORIA HOSP,DEPT DIAGNOST RADIOL,LONDON,ON N6A 4G5,CANADA
[5] UNIV WESTERN ONTARIO,DEPT RADIOL & NUCL MED,LONDON,ON N6A 5C1,CANADA
[6] UNIV WESTERN ONTARIO,DEPT ONCOL,LONDON,ON N6A 5C1,CANADA
[7] UNIV WESTERN ONTARIO,DEPT PHYS,LONDON,ON N6A 5C1,CANADA
关键词
quantum accounting diagram (QAD); detective quantum efficiency (DQE); portal imaging; phosphor screens; amorphous silicon arrays; image quality;
D O I
10.1118/1.598002
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A theoretical approach known as quantum accounting diagram (QAD) analysis has been used to calculate the spatial-frequency-dependent detective quantum efficiency (DQE) of two portal imag ing systems: one based on a video camera and another based on an amorphous silicon array. The spatial frequency-dependent DQEs have then been used to determine indices of displayed and perceived image quality. These indices are figures of merit that can be used to optimize the design of linear imaging systems. We have used this approach to determine which of eight phosphor screen thicknesses (ranging between 67 and 947 mg/cm(2)) is optimal for the two designs of portal imaging systems, The physical characteristics (i.e., detection efficiencies, gains, and MTFs) of each of the eight x-ray detectors have been measured and combined with the physical characteristics of the remaining components to calculate the theoretical DQEs. In turn, the DQEs have been used to calculate theoretical indices of displayed and perceived image quality for two types of objects: a pelvis object and a pointlike object. The maximal indices of displayed and perceived image quality were obtained with screen thicknesses ranging between 358 and 947 mg/cm(2), depending upon the imaging system design and the object being imaged. Importantly, the results showed that there is no single optimal screen thickness. The optimal thickness depended upon imaging task (e.g., detecting large, low-contrast structures, or detecting edges and small structures). Nevertheless, the results showed that there were only modest improvements in the indices of image quality for phosphor screens thicker than 350-400 mg/cm(2). (C) 1997 American Association of Physicists in Medicine.
引用
收藏
页码:803 / 814
页数:12
相关论文
共 40 条
  • [1] ANTONUK LE, 1995, P SOC PHOTO-OPT INS, V2432, P216, DOI 10.1117/12.208340
  • [2] DEMONSTRATION OF MEGAVOLTAGE AND DIAGNOSTIC-X-RAY IMAGING WITH HYDROGENATED AMORPHOUS-SILICON ARRAYS
    ANTONUK, LE
    BOUDRY, J
    HUANG, WD
    MCSHAN, DL
    MORTON, EJ
    YORKSTON, J
    LONGO, MJ
    STREET, RA
    [J]. MEDICAL PHYSICS, 1992, 19 (06) : 1455 - 1466
  • [3] Barrett HH., 1981, RADIOLOGICAL IMAGING
  • [4] CHARACTERIZATION OF FLUORESCENT SCREENS FOR IMAGING APPLICATIONS WITH MEV NEUTRONS AND PHOTONS
    BERZINS, GJ
    LUMPKIN, AH
    SMITH, HL
    [J]. OPTICAL ENGINEERING, 1983, 22 (05) : 633 - 642
  • [5] A quantum accounting and detective quantum efficiency analysis for video-based portal imaging
    Bissonnette, JP
    Cunningham, IA
    Jaffray, DA
    Fenster, A
    Munro, P
    [J]. MEDICAL PHYSICS, 1997, 24 (06) : 815 - 826
  • [6] Evaluation of a high-density scintillating glass for portal imaging
    Bissonnette, JP
    Munro, P
    [J]. MEDICAL PHYSICS, 1996, 23 (03) : 401 - 406
  • [7] OPTIMAL RADIOGRAPHIC MAGNIFICATION FOR PORTAL IMAGING
    BISSONNETTE, JP
    JAFFRAY, DA
    FENSTER, A
    MUNRO, P
    [J]. MEDICAL PHYSICS, 1994, 21 (09) : 1435 - 1445
  • [8] A REVIEW OF ELECTRONIC PORTAL IMAGING DEVICES (EPIDS)
    BOYER, AL
    ANTONUK, L
    FENSTER, A
    VANHERK, M
    MEERTENS, H
    MUNRO, P
    REINSTEIN, LE
    WONG, J
    [J]. MEDICAL PHYSICS, 1992, 19 (01) : 1 - 16
  • [9] EFFICIENCY OF HUMAN VISUAL SIGNAL DISCRIMINATION
    BURGESS, AE
    WAGNER, RF
    JENNINGS, RJ
    BARLOW, HB
    [J]. SCIENCE, 1981, 214 (4516) : 93 - 94
  • [10] A SPATIAL-FREQUENCY DEPENDENT QUANTUM ACCOUNTING DIAGRAM AND DETECTIVE QUANTUM EFFICIENCY MODEL OF SIGNAL AND NOISE-PROPAGATION IN CASCADED IMAGING-SYSTEMS
    CUNNINGHAM, IA
    WESTMORE, MS
    FENSTER, A
    [J]. MEDICAL PHYSICS, 1994, 21 (03) : 417 - 427