Hangman's fracture - A biomechanical comparison of stabilization techniques

被引:95
作者
Duggal, Neil
Chamberlain, Robert H.
Perez-Garza, Luis E.
Espinoza-Larios, Adolfo
Sonntag, Volker K. H.
Crawford, Neil R.
机构
[1] St Joseph Hosp, Barrow Neurol Inst, Phoenix, AZ 85013 USA
[2] Med Ctr, Phoenix, AZ 85013 USA
[3] London Hlth Sci Ctr, London, ON, Canada
[4] Barrow Neurol Inst, Phoenix, AZ USA
关键词
biomechanics; cervical spine; pars; pedicle; hangman'sfracture; spinalinstrumentation;
D O I
10.1097/01.brs.0000251917.83529.0b
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. In vitro biomechanical flexibility experiment studying 5 sequential conditions. Objective. To determine the biomechanical differences among 3 fixation techniques after a simulated hangman's fracture. Summary of Background Data. Type II hangman's fractures are often treated surgically with a C2-C3 anterior cervical discectomy, fusion, and plating. Other techniques include direct fixation with C2 pars interarticularis screws or posterior C2-C3 fixation connecting C2 pars screws to C3 lateral mass screws. Methods. Seven cadaveric specimens (Oc-C4) were tested intact, after a simulated hangman's fracture, and after each fixation technique. Flexion, extension, lateral bending, and axial rotation were induced using nonconstraining torques while recording angular motions stereo-photogrammetrically. Results. Direct screw fixation reduced motion an average of 61% +/- 13% during lateral bending and axial rotation compared to the injured state (P < 0.007). However, instability remained during flexion and extension. Posterior C2-C3 rod fixation provided significantly greater rigidity than anterior plate fixation during lateral bending (P < 0.008) and axial rotation (P < 0.04). Conclusions. Direct fixation of the pars ineffectively limits flexion and extension after a Type II hangman's fracture. If pars screw fixation can be achieved, posterior C2-C3 fixation more effectively stabilizes a hangman's fracture than anterior cervical plating.
引用
收藏
页码:182 / 187
页数:6
相关论文
共 19 条
[1]
TREATMENT OF PEDICULAR FRACTURES OF THE AXIS - A CLINICAL-STUDY AND SCREW FIXATION TECHNIQUE [J].
BORNE, GM ;
BEDOU, GL ;
PINAUDEAU, M .
JOURNAL OF NEUROSURGERY, 1984, 60 (01) :88-93
[2]
Gene expression profiling of pediatric brainstem gliomas [J].
Bristol, RE ;
Rennert, J ;
Coons, S ;
Berens, M .
NEUROSURGERY, 2005, 57 (02) :417-417
[3]
A new technique for determining 3-D joint angles: the tilt/twist method [J].
Crawford, NR ;
Yamaguchi, GT ;
Dickman, CA .
CLINICAL BIOMECHANICS, 1999, 14 (03) :153-165
[4]
AN APPARATUS FOR APPLYING PURE NONCONSTRAINING MOMENTS TO SPINE SEGMENTS IN-VITRO [J].
CRAWFORD, NR ;
BRANTLEY, AGU ;
DICKMAN, CA ;
KOENEMAN, EJ .
SPINE, 1995, 20 (19) :2097-2100
[5]
Construction of local vertebral coordinate systems using a digitizing probe - Technical note [J].
Crawford, NR ;
Dickman, CA .
SPINE, 1997, 22 (05) :559-563
[6]
Crawford NR, 1998, J SPINAL DISORD, V11, P416
[7]
FRACTURES OF THE RING OF THE AXIS - A CLASSIFICATION BASED ON THE ANALYSIS OF 131 CASES [J].
EFFENDI, B ;
ROY, D ;
CORNISH, B ;
DUSSAULT, RG ;
LAURIN, CA .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1981, 63 (03) :319-327
[8]
Craniovertebral junction fixation with transarticular screws: biomechanical analysis of a novel technique [J].
Gonzalez, LF ;
Crawford, NR ;
Chamberlain, RH ;
Garza, LEP ;
Preul, MC ;
Sonntag, VKH ;
Dickman, CA .
JOURNAL OF NEUROSURGERY, 2003, 98 (02) :202-209
[9]
Biomechanical evaluation of occipitocervical instrumentation: screw compared with wire fixation [J].
Hurlbert, RJ ;
Crawford, NR ;
Choi, WG ;
Dickman, CA .
JOURNAL OF NEUROSURGERY, 1999, 90 (01) :84-90
[10]
KARIMI-NEJAD A, 1989, Neurosurgical Review, V12, P525, DOI 10.1007/BF01790699