Reduced ventilator pressure and improved P/F ratio during percutaneous arteriovenous carbon dioxide removal for severe respiratory failure

被引:31
作者
Alpard, SK
Zwischenberger, JB [1 ]
Tao, WK
Deyo, DJ
Bidani, A
机构
[1] Univ Texas, Med Branch, Dept Surg, Galveston, TX 77555 USA
[2] Univ Texas, Med Branch, Dept Med, Galveston, TX 77555 USA
[3] Univ Texas, Med Branch, Dept Anesthesiol, Galveston, TX 77555 USA
[4] Shriners Burns Inst, Galveston, TX USA
关键词
D O I
10.1097/00000658-199908000-00012
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objective To evaluate the effect of percutaneous arteriovenous carbon dioxide removal (AVCO(2)R) on ventilator pressures and P/F ratio in a clinically relevant large-animal model of severe respiratory failure. Summary Background Data AVCO(2)R was developed as a simple arteriovenous shunt with a commercially available low-resistance gas exchange device of sufficient surface area for near-total CO2 removal. With an AV shunt 10% to 15% of cardiac output, AVCO(2)R allows a reduction in ventilator airway pressures without hypercapnia or the complex circuitry and monitoring required for conventional ECMO. Methods AVCO(2)R was applied to a new, clinically relevant large-animal model of severe respiratory failure created by smoke inhalation and cutaneous flame burn injury. Adult sheep (n = 9, 38 +/- 6 kg) received a 40% total body surface area, third-degree cutaneous flame burn and 36 breaths of cotton smoke insufflation. After injury, all animals were placed on volume-controlled mechanical ventilation to achieve PaO2 > 60 mmHg and PaCO2 < 40 mmHg. Animals were placed AVCO(2)R within 40 to 48 hours of injury when the PaO2/FiO(2), was <200. Animals underwent cannulation of the carotid artery and jugular vein with percutaneous 10F arterial and 14F venous cannulas. Shunt flow was continuously monitored using an ultrasonic flow probe and calculated as a percentage of cardiac output. Results AVCO(2)R flows of 800 to 900 ml/min (11% to 13% cardiac output) achieved 77 to 104 ml/min of CO2 removal (95% to 97% total CO2 production) while maintaining normocapnia. Significant reductions in ventilator settings were tidal volume, 421.3 +/- 39.8 to 270.0 +/- 6.3 ml; peak inspiratory pressure, 24.8 +/- 2.4 to 13.7 +/- 0.7 cm H2O; minute ventilation, 12.7 +/- 1.4 to 6.2 +/- 0.8 L/min; respiratory rate, 25.4 +/- 1.3 to 18.4 +/- 1.8 breaths/min; and FiO(2), 0.88 +/- 0.1 to 0.39 +/- 0.1. The P/F ratio increased from 151.5 +/- 40.0 at baseline to 320.0 +/- 17.8 after 72 hours. Conclusions Percutaneous AVCO(2)R allows near-total CO2 removal and significant reductions in ventilator pressures with improvement in the P/F ratio.
引用
收藏
页码:215 / 224
页数:10
相关论文
共 61 条
[31]  
GRIFFIN MP, 1991, CIRC SHOCK, V33, P195
[32]   EVIDENCE OF LUNG SURFACTANT ABNORMALITY IN RESPIRATORY-FAILURE - STUDY OF BRONCHOALVEOLAR LAVAGE PHOSPHOLIPIDS, SURFACE-ACTIVITY, PHOSPHOLIPASE-ACTIVITY, AND PLASMA MYOINOSITOL [J].
HALLMAN, M ;
SPRAGG, R ;
HARRELL, JH ;
MOSER, KM ;
GLUCK, L .
JOURNAL OF CLINICAL INVESTIGATION, 1982, 70 (03) :673-683
[33]   INHALATION INJURY IN BURNS [J].
HEAD, JM .
AMERICAN JOURNAL OF SURGERY, 1980, 139 (04) :508-512
[34]  
Herndon D N, 1985, Crit Care Clin, V1, P79
[35]   VENTILATORY MANAGEMENT OF ARDS - CAN IT AFFECT THE OUTCOME [J].
HICKLING, KG .
INTENSIVE CARE MEDICINE, 1990, 16 (04) :219-226
[36]   ELEVATED LEVELS OF ENDOTOXIN, OXYGEN-DERIVED FREE-RADICALS, AND CYTOKINES DURING EXTRACORPOREAL MEMBRANE-OXYGENATION [J].
HIRTHLER, M ;
SIMONI, J ;
DICKSON, M .
JOURNAL OF PEDIATRIC SURGERY, 1992, 27 (09) :1199-1202
[37]  
HOCKER JR, 1991, ARTIF ORGANS, V15, P23
[38]  
Ichiba S, 1996, ARTIF ORGANS, V20, P120
[39]   INCREASING DURATION OF SMOKE EXPOSURE INDUCES MORE SEVERE LUNG INJURY IN SHEEP [J].
KIMURA, R ;
TRABER, LD ;
HERNDON, DN ;
LINARES, HA ;
LUBBESMEYER, HJ ;
TRABER, DL .
JOURNAL OF APPLIED PHYSIOLOGY, 1988, 64 (03) :1107-1113
[40]   Extracorporeal life support for 100 adult patients with severe respiratory failure [J].
Kolla, S ;
Awad, SS ;
Rich, PB ;
Schreiner, RJ ;
Hirschl, RB ;
Bartlett, RH .
ANNALS OF SURGERY, 1997, 226 (04) :544-564