End-tidal PCO2 monitoring via nasal cannulae in pediatric patients: Accuracy and sources of error

被引:38
作者
Friesen, RH [1 ]
Alswang, M [1 ]
机构
[1] CHILDRENS HOSP,DEPT ANESTHESIOL & PEDIAT CRIT CARE,DENVER,CO 80218
来源
JOURNAL OF CLINICAL MONITORING | 1996年 / 12卷 / 02期
关键词
measurement techniques; capnography; Monitoring; carbon dioxide; blood gases; nasal cannulae; pediatrics;
D O I
10.1007/BF02078136
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Objective. To assess the correlation and accuracy of end-tidal PCO2 (PetCO(2)) sampled via nasal cannulae in pediatric patients by comparison to the criterion standard PaCO2 and to identify sources of error during PetCO(2) monitoring via nasal cannulae. Methods. PetCO(2) was monitored continuously by sampling end-tidal gas through nasal cannulae that had been designed and manufactured for this purpose in spontaneously breathing children undergoing conscious or deep sedation during either cardiac catheterization (n = 43) or critical care (n = 54). When both the capnographic wave form and the PetCO(2) value had been stable for at least 10 minutes, the PetCO(2) value was recorded while blood was drawn from atl indwelling arterial line for PaCO2 measurement. The effects of age, weight, respiratory rate, oxygen delivery system, airway obstruction, mouth breathing, and cyanotic heart disease were evaluated by linear regression analysis and calculation of absolute bias (PaCO2-PetCO(2)). Results. Mouth breathing, airway obstruction, oxygen delivery through the ipsilateral nasal cannula, and cyanotic heart disease adversely affected accuracy. In patients without those factors, PetCO(2) correlated well with PaCO2 (R(2) = 0.994), and absolute bias was 3.0 +/- 1.8 mmHg. Conclusions. Several factors - some controllable and all recognizable - affect the accuracy of PetCO(2) monitored via nasal cannulae in pediatric patients. When these factors are not present, PetCO(2) correlates well with PaCO2 and appears to be a useful monitor of ventilatory status during conscious or deep sedation.
引用
收藏
页码:155 / 159
页数:5
相关论文
共 19 条
[1]  
Ackerman W E, 1990, Anesth Prog, V37, P199
[2]  
BADGWELL JM, 1987, ANESTH ANALG, V66, P959
[3]   FREQUENT HYPOXEMIA AND APNEA AFTER SEDATION WITH MIDAZOLAM AND FENTANYL [J].
BAILEY, PL ;
PACE, NL ;
ASHBURN, MA ;
MOLL, JWB ;
EAST, KA ;
STANLEY, TH .
ANESTHESIOLOGY, 1990, 73 (05) :826-830
[4]   ACCURATE DETERMINATION OF END-TIDAL CARBON-DIOXIDE DURING ADMINISTRATION OF OXYGEN BY NASAL CANNULAE [J].
BOWE, EA ;
BOYSEN, PG ;
BROOME, JA ;
KLEIN, EF .
JOURNAL OF CLINICAL MONITORING, 1989, 5 (02) :105-110
[5]   Changes in carbon dioxide tension and oxygen saturation during sleep sedation for paediatric cardiac catheterization [J].
Friesen, RH ;
Alswang, M .
PAEDIATRIC ANAESTHESIA, 1996, 6 (01) :15-20
[6]   A SIMPLE, EASY, AND INEXPENSIVE METHOD FOR MONITORING ETCO2 THROUGH NASAL CANNULAE [J].
GOLDMAN, JM .
ANESTHESIOLOGY, 1987, 67 (04) :606-606
[7]  
Gravenstein N, 1985, J Clin Monit, V1, P6, DOI 10.1007/BF02832683
[8]  
Hunter J A, 1990, Anesth Prog, V37, P301
[9]   RELATIONSHIP BETWEEN INVASIVE AND NONINVASIVE MEASUREMENTS OF GAS-EXCHANGE IN ANESTHETIZED INFANTS AND CHILDREN [J].
LINDAHL, SGE ;
YATES, AP ;
HATCH, DJ .
ANESTHESIOLOGY, 1987, 66 (02) :168-175
[10]   RELATIONSHIP BETWEEN ARTERIAL CARBON-DIOXIDE AND END-TIDAL CARBON-DIOXIDE WHEN A NASAL SAMPLING PORT IS USED [J].
MCNULTY, SE ;
ROY, J ;
TORJMAN, M ;
SELTZER, JL .
JOURNAL OF CLINICAL MONITORING, 1990, 6 (02) :93-98