Development of a polymer-based easy-to-fabricate micro-free-flow electrophoresis device

被引:6
作者
Akagi, Takanori [1 ]
Kubota, Ryosuke [1 ]
Kobayashi, Masashi [1 ]
Ichiki, Takanori [1 ]
机构
[1] Univ Tokyo, Sch Engn, Dept Bioengn, Bunkyo Ku, Tokyo 1138656, Japan
基金
日本科学技术振兴机构;
关键词
SAMPLE PURIFICATION; MASS-SPECTROMETRY; SEPARATION; HYDROGEL; PROTEINS; AFFINITY; SCALE; GELS; IEF;
D O I
10.7567/JJAP.54.06FN05
中图分类号
O59 [应用物理学];
学科分类号
摘要
Since 1990s, micro-free-flow electrophoresis (mu FFE) devices have been developed to allow for smaller sample volume and reagent consumption. To solve several technical problems involving the generation of electrolysis gas on the electrodes, most of the mu FFE devices reported in the past were fabricated using elaborate micromachining process on silicon or glass substrates. However, high-cost micromachining processes were required and these were not suitable for mass production. In this paper, we report a polymer-based easy-to-fabricate mu FFE device using a poly(methyl methacrylate-co-styrene), P(MMA-co-S), substrate and tetra-PEG gel for preventing the invasion of electrolysis gas into the separation chamber. In the separation experiment using a mixture of rhodamine B and sulforhodamine B, the resolution increased linearly with the increase of the applied voltages up to 50V, whereas a deviation from the linear relation was observed above 50V, which is possibly the Joule heating. These results indicate that this device could be applicable to separation of biological samples. (C) 2015 The Japan Society of Applied Physics
引用
收藏
页数:5
相关论文
共 32 条
[1]   Micro free-flow IEF enhanced by active cooling and functionalized gels [J].
Albrecht, Jacob W. ;
Jensen, Klavs F. .
ELECTROPHORESIS, 2006, 27 (24) :4960-4969
[2]   Microchip free-flow electrophoresis on glass substrate using laser-printing toner as structural material [J].
de Jesus, Dosil Pereira ;
Blanes, Lucas ;
do Lago, Claudimir Lucio .
ELECTROPHORESIS, 2006, 27 (24) :4935-4942
[3]   Micro total analysis systems. Latest advancements and trends [J].
Dittrich, Petra S. ;
Tachikawa, Kaoru ;
Manz, Andreas .
ANALYTICAL CHEMISTRY, 2006, 78 (12) :3887-3907
[4]   Using channel depth to isolate and control flow in a micro free-flow electrophoresis device [J].
Fonslow, Bryan R. ;
Barocas, Victor H. ;
Bowser, Michael T. .
ANALYTICAL CHEMISTRY, 2006, 78 (15) :5369-5374
[5]   Optimizing band width and resolution in micro-free flow electrophoresis [J].
Fonslow, Bryan R. ;
Bowser, Michael T. .
ANALYTICAL CHEMISTRY, 2006, 78 (24) :8236-8244
[6]  
Gobom J, 1999, J MASS SPECTROM, V34, P105, DOI 10.1002/(SICI)1096-9888(199902)34:2<105::AID-JMS768>3.3.CO
[7]  
2-W
[8]   DIE TRAGERFREIE KONTINUIERLICHE ELEKTROPHORESE UND IHRE ANWENDUNG [J].
HANNIG, K .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1961, 181 :244-274
[9]   ANALYSIS OF ACTIVE ANTIBODY CONCENTRATION - SEPARATION OF AFFINITY AND CONCENTRATION PARAMETERS [J].
KARLSSON, R ;
FAGERSTAM, L ;
NILSHANS, H ;
PERSSON, B .
JOURNAL OF IMMUNOLOGICAL METHODS, 1993, 166 (01) :75-84
[10]   Effect of demulsifier partitioning on the destabilization of water-in-oil emulsions [J].
Kim, YH ;
Wasan, DT .
INDUSTRIAL & ENGINEERING CHEMISTRY RESEARCH, 1996, 35 (04) :1141-1149