Monte Carlo calculations and GafChromic film measurements for plugged collimator helmets of Leksell Gamma Knife unit

被引:21
作者
Cheung, JYC [1 ]
Yu, KN
Ho, RTK
Yu, CP
机构
[1] Canossa Hosp, Gamma Knife Ctr HK, 1 Old Peak Rd, Hong Kong, Peoples R China
[2] City Univ Hong Kong, Dept Phys & Mat Sci, Kowloon, Peoples R China
关键词
Gamma Knife; GammaPlan; plug pattern; Monte Carlo; GafChromic film;
D O I
10.1118/1.598639
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
The Monte Carlo technique and GafChromic films were employed to verify the accuracy of the dose planning system (Leksell GammaPlan(R)) used in Gamma Knife (type B) radiosurgery when plugged collimator helmets were used. The EGS4 Monte Carlo code was used to calculate the dose distribution along the x, y, and z axes when a single shot was delivered at the center point (unit center point: x=100, y=100, z=100) of a spherical polystyrene phantom, with gamma angle of 90 degrees. Two different sizes of the plugged collimator helmets, 4 and. 18 mm, were studied. Two typical. plugged patterns, 51 plugs and 99 plugs along the y direction, were: examined. The results of our Monte Carlo trials showed good consistency with GammaPlan calculations and GafChromic sim measurements. Furthermore, the Monte Carlo results showed that radiation leakage from the plugs was too small to affect the overall isodose curve distribution even when the heavily plugged pattern of up to 99 plugs was employed. The results of this project provide confidence to all Gamma Knife centers using the Leksell GammaPlan treatment planning system. (C) 1999 American Association of Physicists in Medicine. [S0094-2405(99)02207-5].
引用
收藏
页码:1252 / 1256
页数:5
相关论文
共 7 条
[1]
*AM, 1982, MED RAD SOURC CAT
[2]
*EL, 1991, LEKS GAMM PLAN INSTR
[3]
*EL, 1992, LEKS GAMM UN US MAN, V1
[4]
Hartmann G, 1995, QUALITY ASSURANCE PR
[5]
Jenkins T M., 1988, Monte Carlo Transport of Electrons and Photons
[6]
PHYSICS OF GAMMA-KNIFE APPROACH ON CONVERGENT BEAMS IN STEREOTAXIC RADIOSURGERY [J].
WU, A ;
LINDNER, G ;
MAITZ, AH ;
KALEND, AM ;
LUNSFORD, LD ;
FLICKINGER, JC ;
BLOOMER, WD .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 18 (04) :941-949
[7]
Wu A, 1992, Neurosurg Clin N Am, V3, P35