Volar fixation for dorsally angulated extra-articular fractures of the distal radius: A biomechanical study

被引:69
作者
Koh, Shukuki
Morris, Randal P.
Patterson, Rita A.
Kearney, J. Patrick
Buford, William L., Jr.
Viegas, Steven F.
机构
[1] Univ Texas, Med Branch, Dept Orthopaed Surg & Rehabil, Div Hand Surg, Galveston, TX 77555 USA
[2] Nagoya Univ, Sch Med, Dept Hand Surg, Nagoya, Aichi 466, Japan
来源
JOURNAL OF HAND SURGERY-AMERICAN VOLUME | 2006年 / 31A卷 / 05期
关键词
biomechanics; distal radius fracture; internal fixation; plate;
D O I
10.1016/j.jhsa.2006.02.015
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose: To compare the biomechanical properties of 10 volar plate-fixation designs in 2 fracture models (dorsal wedge osteotomy, segmental resection osteotomy models). Methods: Forty-eight radiuses were used in this study including 8 pairs. In 40 specimens a 15-mm dorsally based wedge osteotomy was performed and the volar cortex was fractured manually. They were arranged into 10 fixation groups with 5 different fixation designs (test 1). In the contralateral specimens of 8 paired radiuses a 10-mm segment of bone was excised (test 2). Four of the 10 fixation systems were chosen for these specimens. Cadaver hands and the proximal radiuses were potted in polymethylmethacrylate and tested with a servohydraulic materials testing machine with 300 N of axial compression load at 1 N/s initially and after each 1,000 cycles up to 5,000 cycles. After cyclic loading the specimens were loaded to failure in axial compression at 2 mm/min. The stiffness, failure peak load, and failure mode of each specimen were recorded. Results: In test 1 in the wedge osteotomy specimens the T plate was the stiffest and the Synthes titanium plate was the least stiff; however, all specimens completed the 5,000 cycles of loading with no failures. There was no significant difference between the 10 fixation groups in failure peak load and only 7 of 40 failed at the distal portion of the hardware in the final load to failure testing. In test 2 the resection osteotomy specimens were less stiff and failed at a lower failure peak load compared with the wedge osteotomy specimens. Failure at the distal portion of the fixation system was seen in 7 of 8 specimens; nonlocking screws loosened and tines compressed the surrounding bone, resulting in tine-hole enlargement. Conclusions: All of the plate-fixation systems delivered sufficient stability to permit the simulated postoperative regimen of I week of immobilization followed by 5 weeks of early mobilization until expected union at 6 weeks after surgery. Based on these results a preferable volar fixation system would appear to benefit from the following: (1) sufficient plate strength to support the distal fragment from the volar side, (2) a locking system with sufficient strength to remain locked during the healing process, and (3) a distal design that does not affect the bone adversely. The anatomic reduction of the volar cortex in the wedge osteotomy specimens added stability to the construct.
引用
收藏
页码:771 / 779
页数:9
相关论文
共 17 条
[1]
Open reduction and internal fixation of unstable distal radius fractures with a low-profile plate: A multicenter study of 73 fractures [J].
Carter, PR ;
Frederick, HA ;
Laseter, GF .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 1998, 23A (02) :300-307
[2]
Radial artery pseudoaneurysm complication from use of AO/ASIF volar distal radius plate: A case report [J].
Dao, KD ;
Venn-Watson, E ;
Shin, AY .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2001, 26A (03) :448-453
[3]
FERNANDEZ DL, 2002, FRACTURES DISTAL RAD, P42
[4]
FERNANDEZ DL, 2002, FRACTURES DISTAL RAD, P131
[5]
GREEN DP, 2005, GREENS OPERATIVE HAN, P680
[6]
COMPRESSION STRENGTH OF HUMAN RADII [J].
HOYER, HE ;
KOKEMOHR, M ;
LIPPERT, H .
ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 1980, 118 (05) :816-821
[7]
JENKINS NH, 1988, J BONE JOINT SURG, V70, P291
[8]
Complications of the AO/ASIF titanium distal radius plate system (π plate) in internal fixation of the distal radius:: A brief report [J].
Kambouroglou, GK ;
Axelrod, TS .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 1998, 23A (04) :737-741
[9]
COLLES FRACTURE - DOES THE ANATOMICAL RESULT AFFECT THE FINAL FUNCTION [J].
MCQUEEN, M ;
CASPERS, J .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1988, 70 (04) :649-651
[10]
Delayed rupture of the flexor pollicis longus tendon after inappropriate placement of the π plate on the volar surface of the distal radius [J].
Nunley, JA ;
Rowan, PR .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 1999, 24A (06) :1279-1280