共 6 条
Thoracic spiral CT: Delivery of contrast material pushed with injectable saline solution in a power injector
被引:99
作者:
Hopper, KD
[1
]
Mosher, TJ
[1
]
Kasales, CJ
[1
]
TenHave, TR
[1
]
Tully, DA
[1
]
Weaver, JS
[1
]
机构:
[1] PENN STATE UNIV,MILTON S HERSHEY MED CTR,CTR BIOSTAT & EPIDEMIOL,HERSHEY,PA 17033
来源:
关键词:
computed tomography (CT);
index media;
helical;
contrast media;
experimental studies;
thorax;
CT;
D O I:
10.1148/radiology.205.1.9314997
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
A method of power injection of contrast material pushed with injectable saline solution during thoracic computed tomography (CT) was evaluated in 75 patients to help decrease the amount of contrast material necessary. Patients received 50-125 ml of 60% nonionic contrast material alone or pushed with 50 ml of saline. A volume of 75 ml of contrast material pushed with saline provided enhancement of the thoracic vasculature equal to that provided by a 125-mL volume of contrast material administered alone (P <.001) and caused significantly less beam-hardening artifact (P < .01).
引用
收藏
页码:269 / 271
页数:3
相关论文