Novel instrumentation for real-time monitoring using miniaturized flow systems with integrated biosensors

被引:33
作者
Freaney, R
McShane, A
Keaveny, TV
McKenna, M
Rabenstein, K
Scheller, FW
Pfeiffer, D
Urban, G
Moser, I
Jobst, G
Manz, A
Verpoorte, E
Widmer, MW
Diamond, D
Dempsey, E
deViteri, FJS
Smyth, M
机构
[1] MAX DELBRUCK INST,BERLIN,GERMANY
[2] VIENNA TECH UNIV,VIENNA,AUSTRIA
[3] CIBA GEIGY AG,CH-4002 BASEL,SWITZERLAND
[4] UNIV LONDON IMPERIAL COLL SCI TECHNOL & MED,DEPT CHEM,LONDON SW7 2AY,ENGLAND
[5] DUBLIN CITY UNIV,DUBLIN 9,IRELAND
关键词
monitoring; physiologic; biosensor; microdialysis; blood glucose; lactate;
D O I
10.1177/000456329703400312
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
A prototype miniaturized Total Chemical Analysis System (mu TAS) has been developed and applied to on-line monitoring of glucose and lactate in the core blood of anaesthetized dogs. The system consists of a highly efficient microdialysis sampling interface sited in a small-scale extracorporeal shunt circuit ('MiniShunt'), a silicon machined microflow manifold and integrated biosensor array for glucose and lactate detection with associated computer software for analytical process control. During in-vivo testing the device allowed real-time on-screen monitoring of glucose and lactate with system response times of less than 5 min, made possible by the small dead volume of the microflow system. On-line glucose and lactate measurements were made in the basal state as well as during intravenous infusion of glucose or lactate. The prototype mu TAS is currently suitable for trend monitoring but refinements are necessary before application of the system for determination of individual lactate values.
引用
收藏
页码:291 / 302
页数:12
相关论文
共 49 条
[1]  
ALBERTI TJ, 1990, CRIT CARE MED, V18, P275
[2]  
ALCOCK SJ, 1994, IEEE ENG MED BIO JUN, P319
[3]   APPLICATION OF CHRONIC INTRAVASCULAR BLOOD-GLUCOSE SENSOR IN DOGS [J].
ARMOUR, JC ;
LUCISANO, JY ;
MCKEAN, BD ;
GOUGH, DA .
DIABETES, 1990, 39 (12) :1519-1526
[4]   MICRODIALYSIS OF ADIPOSE-TISSUE [J].
ARNER, P ;
BOLINDER, J .
JOURNAL OF INTERNAL MEDICINE, 1991, 230 (04) :381-386
[5]   MICRODIALYSIS - THEORY AND APPLICATION [J].
BENVENISTE, H ;
HUTTEMEIER, PC .
PROGRESS IN NEUROBIOLOGY, 1990, 35 (03) :195-215
[6]   PROGRESS IN PRACTICAL ENDOCRINOLOGY - THE GLUCOSENSOR UNITEC ULM - A PORTABLE MONITOR FOR CONTINUOUS BLOOD-GLUCOSE MEASUREMENT [J].
BRUCKEL, J ;
ZIER, H ;
KERNER, W ;
PFEIFFER, EF .
HORMONE AND METABOLIC RESEARCH, 1990, 22 (07) :382-384
[7]  
DEBOER J, 1991, PFLUG ARCH EUR J PHY, V419, P1
[8]  
DEVITERI FJS, 1994, ANAL PROC, V31, P229
[9]  
*DIAB CONTR COMPL, 1993, NEW ENGL J MED, V329, P977, DOI DOI 10.1056/NEJM199309303291401
[10]   CLINICAL USEFULNESS OF THE GLUCOSE-CONCENTRATION IN THE SUBCUTANEOUS TISSUE - PROPERTIES AND PITFALLS OF ELECTROCHEMICAL BIOSENSORS [J].
FISCHER, U ;
REBRIN, K ;
VONWOEDTKE, T ;
ABEL, P .
HORMONE AND METABOLIC RESEARCH, 1994, 26 (11) :515-522