Novel, side opened out-of-plane microneedles for microfluidic transdermal interfacing

被引:43
作者
Griss, P [1 ]
Stemme, G [1 ]
机构
[1] Royal Inst Technol, Dept Signals Sensors & Syst, S-10044 Stockholm, Sweden
来源
FIFTEENTH IEEE INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS, TECHNICAL DIGEST | 2002年
关键词
D O I
10.1109/MEMSYS.2002.984303
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We present the first hollow out of wafer plane silicon microneedles that have openings in the shaft rather than having an orifice at the tip. These structures are suited for transdermal microfluidic applications, e.g. drug- or vaccine delivery. The developed deep reactive ion etching (DRIE) process allows fabrication of two dimensional, mechanically highly resistant needle arrays offering low resistance to liquid flows and offering a large exposure area between the fluid and the tissue. The presented process does not require much wafer handling and only two photolithography steps are needed. Using this chip in a typical application, e.g. vaccine delivery, a 100 mul volume of aqueous fluid injected through a chip of 30 mm(2) in 2s would cause a pressure drop of less than 2kPa. The presented needles are approximately 2 10 pm long.
引用
收藏
页码:467 / 470
页数:4
相关论文
共 13 条
[11]  
Stoeber B., 2000, 1 ANN INT IEEE EMBS
[12]  
STUPAR PA, 2001, 11 INT C SOL STAT SE
[13]  
TALBOT NH, 1998, POL 2 WAF POL MICR C