Precise CD4 T-cell counting using red diode laser excitation: For richer, for poorer

被引:37
作者
Janossy, G
Jani, IV
Kahan, M
Barnett, D
Mandy, F
Shapiro, H
机构
[1] UCL Royal Free & Univ Coll Med Sch, Dept Immunol & Mol Pathol, London NW3 2PF, England
[2] Hlth Canada, Ctr Infect Dis Prevent & Control, Natl lab HIV Immunol, Ottawa, ON K1A 0L2, Canada
来源
CYTOMETRY | 2002年 / 50卷 / 02期
关键词
flow cytometry; red diode lasers; leucocyte differentials; immunophenotyping; CD4 T lymphocytes; absolute CD4 counts; primary CD4 gating; pan-leucogating; CD4% values; human immunodeficiency virus (HIV); AIDS; morphospectral attributes; developing countries;
D O I
10.1002/cyto.10082
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Background: Measuring CD4 T-cell counts at low cost is relevant in dealing with the human immunodeficiency virus (HIV) epidemic throughout the developing world. The recently introduced novel concepts in gating strategies and sample stabilization facilitate affordable immunophenotyping by flow cytometry. However, the impact of these developments is still limited by the high cost of currently available flow cytometers. Methods: Diode lasers emitting 10-15 mW at 635 nm are one-tenth the size and cost and require one thousandth the power of an equivalent 488-nm argon ion laser. We used the available 635-nm diode-based flow cytometers, including PA-II, Luminex 100, SuperMot, and FACSCalibur, to investigate whether these instruments can generate reliable CD4 counts when used with allophycocyanin (APC) and cyanin-5 (Cy5)-labeled CD4 antibodies. Results: We document the feasibility of obtaining leucocyte differential counts using orthogonal side scatter (SSC) without the need for forward scatter (FSC). Accurate CD4% values among lymphocytes and leucocytes can be obtained by primary CD4 gating using a single CD4 monoclonal antibody conjugated to APC or Cy5. Double immunofluorescence (IF) staining with CD4-APC (FL1) and CD45-APC-Cy7 (FL2) introduces pan-leucogating for a convenient assessment of absolute CD4 counts on double platforms. We demonstrate that small flow cytometers with laser diodes are capable of delivering absolute CD4 T-cell counts with a precision similar to the performance of the current state-of-the-art single-platform instruments (e.g., the CytoronAbsolut&, R-2 = 0.961). In this respect, they appear to be superior to the nonflow CD4 counting techniques. Conclusions: Accurate CD4 counts can he generated at minimal cost on red diode laser-operated flow cytometers, retaining the potential for high throughput capacity without compromising precision. With further improvements in volumetric technology and clinical software, these cytometers may develop into a new generation of inexpensive battery-operated laboratory hardware that combines cellular phenotyping with head-based multiplexing immunoassays for (HIV) serology. Cytometry (Clin. Cytometry) 50:78-85, 2002. (C) 2002 Wiley-Liss, Inc.
引用
收藏
页码:78 / 85
页数:8
相关论文
共 37 条
  • [1] Barnett D, 1996, CYTOMETRY, V26, P216, DOI 10.1002/(SICI)1097-0320(19960915)26:3<216::AID-CYTO6>3.0.CO
  • [2] 2-C
  • [3] BARNETT D, IN PRESS BR J HAEMAT
  • [4] Selection of lymphocyte gating protocol has an impact on the level of reliability of T-cell subsets in aging specimens
    Bergeron, M
    Nicholson, JKA
    Phaneuf, S
    Ding, T
    Soucy, N
    Badley, AD
    Foss, NCH
    Mandy, F
    [J]. CYTOMETRY, 2002, 50 (02): : 53 - 61
  • [5] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [6] Corver WE, 2000, CYTOMETRY, V39, P96, DOI 10.1002/(SICI)1097-0320(20000201)39:2<96::AID-CYTO2>3.0.CO
  • [7] 2-X
  • [8] 11-color, 13-parameter flow cytometry: Identification of human naive T cells by phenotype, function, and T-cell receptor diversity
    De Rosa, SC
    Herzenberg, LA
    Herzenberg, LA
    Roederer, M
    [J]. NATURE MEDICINE, 2001, 7 (02) : 245 - 248
  • [9] Comparative assessment of five alternative methods for CD4+ T-lymphocyte enumeration for implementation in developing countries
    Didier, JM
    Kazatchkine, MD
    Demouchy, C
    Moat, C
    Diagbouga, S
    Sepulveda, C
    Di Lonardo, AM
    Weiss, L
    [J]. JOURNAL OF ACQUIRED IMMUNE DEFICIENCY SYNDROMES, 2001, 26 (02) : 193 - 195
  • [10] VISIBLE DIODE-LASERS CAN BE USED FOR FLOW CYTOMETRIC IMMUNOFLUORESCENCE AND DNA ANALYSIS
    DOORNBOS, RMP
    DEGROOTH, BG
    KRAAN, YM
    VANDERPOEL, CJ
    GREVE, J
    [J]. CYTOMETRY, 1994, 15 (03): : 267 - 271