Inhibition of bradycardia pacing caused by far-field atrial sensing in a third-generation cardioverter defibrillator with an automatic gain feature

被引:16
作者
Curwin, JH [1 ]
Roelke, M [1 ]
Ruskin, JN [1 ]
机构
[1] MASSACHUSETTS GEN HOSP,CARDIA ARRHYTHMIA SERV,CARDIAC UNIT,BOSTON,MA 02114
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1996年 / 19卷 / 01期
关键词
implantable cardioverter defibrillator; automatic gain; asystole;
D O I
10.1111/j.1540-8159.1996.tb04802.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The diagnostic accuracy of implantable cardioverter defibrillators may be improved by automatically adjusting gain algorithms, which in general reduce the likelihood of oversensing while maintaining the ability to detect the low amplitude signals associated with ventricular fibrillation. We present a patient with a third-generation device who developed prolonged ventricular asystole arising as a complication of the automatic gain feature. During asystole the device automatically increased sensitivity in order to prevent undersensing of ventricular fibrillation, which in this case resulted in far-field sensing of atrial activity and inhibition of ventricular pacing.
引用
收藏
页码:124 / 126
页数:3
相关论文
共 7 条
[1]   PACED BEATS FOLLOWING SINGLE NONSENSED COMPLEXES IN A CODEPENDENT CARDIOVERTER DEFIBRILLATOR AND BRADYCARDIA PACING SYSTEM - POTENTIAL FOR VENTRICULAR-TACHYCARDIA INDUCTION [J].
CALLANS, DJ ;
HOOK, BG ;
MARCHLINSKI, FE .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (08) :1281-1287
[2]   ELECTROCARDIOGRAPHICALLY DOCUMENTED UNNECESSARY, SPONTANEOUS SHOCKS IN 241 PATIENTS WITH IMPLANTABLE CARDIOVERTER DEFIBRILLATORS [J].
GRIMM, W ;
FLORES, BF ;
MARCHLINSKI, FE .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (11) :1667-1673
[3]   TACHYCARDIA SENSING FAILURE OF AN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR IN A PATIENT WITH HYPERTROPHIC CARDIOMYOPATHY [J].
GRUBB, BP ;
DURZINSKY, D ;
TEMESYARMOS, P ;
HAHN, H ;
ELLIOTT, L .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (06) :845-848
[4]   IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR THERAPY IN THE ABSENCE OF SIGNIFICANT SYMPTOMS - RHYTHM DIAGNOSIS AND MANAGEMENT AIDED BY STORED ELECTROGRAM ANALYSIS [J].
HOOK, BG ;
CALLANS, DJ ;
KLEIMAN, RB ;
FLORES, BT ;
MARCHLINSKI, FE .
CIRCULATION, 1993, 87 (06) :1897-1906
[5]   CONSIDERATIONS FOR VENTRICULAR-FIBRILLATION DETECTION BY IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS [J].
JONES, GK ;
BARDY, GH .
AMERICAN HEART JOURNAL, 1994, 127 (04) :1107-1110
[6]   OVERSENSING DURING VENTRICULAR PACING IN PATIENTS WITH A 3RD-GENERATION IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR [J].
KELLY, PA ;
MANN, DE ;
DAMLE, RS ;
REITER, MJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 23 (07) :1531-1534
[7]   FAILURE OF A 2ND AND 3RD GENERATION IMPLANTABLE CARDIOVERTER DEFIBRILLATOR TO SENSE VENTRICULAR-TACHYCARDIA - IMPLICATIONS FOR FIXED-GAIN SENSING DEVICES [J].
SPERRY, RE ;
ELLENBOGEN, KA ;
WOOD, MA ;
STAMBLER, BS ;
DIMARCO, JP ;
HAINES, DE .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (05) :749-755